Real-time ultrasound and freehand-SPECT. Experiences with sentinel lymph node mapping.
Freesmeyer, M; Winkens, T; Opfermann, T; Elsner, P; Runnebaum, I; Darr, A.
Afiliação
  • Freesmeyer M; Martin Freesmeyer, M.D., Clinic of Nuclear Medicine, Jena University Hospital, Bachstraße 18, 07743 Jena, Germany, Tel. +49/(0)36 41/93 32 20, Fax +49/(0)36 41/93 32 44, E-mail: martin.freesmeyer@med.uni-jena.de.
Nuklearmedizin ; 53(6): 259-64, 2014.
Article em En | MEDLINE | ID: mdl-25056555
ABSTRACT
UNLABELLED Aim of this work is to report first experiences of the feasibility and applicability of a hybrid freehandSPECT/ultrasound (fh-SPECT/US) imaging concept, with regard to SLN imaging, in patients with breast cancer and malignant melanoma. PATIENTS,

METHODS:

18 patients with breast cancer or malignant melanoma received standard SLN scintigraphy. Following this, fh-SPECT using declipse®SPECT (SurgicEye, Munich, Germany) was performed, a handheld-gamma camera-based method to visualize activity distribution within a region of interest as a cross-sectional data set. These data were transferred to an ultrasound device and sensor-navigated ultrasound was performed combining fh-SPECT data with ultrasound images, displaying superimposed images. Quality of fh-SPECT and co-registration accuracy was assigned to one of four categories and occurrence of artefacts was assessed.

RESULTS:

In 4/18 examinations, there was a no deviation regarding co-registration of both data sets. For 9/18 patients, there was a deviation of <1 cm (mean 0.7±0.3 cm, range 0.3-1.0 cm). For 3/18 patients, a deviation >1 cm was present (mean 1.7±0.3 cm, range 1.5-2.0 cm). In 2/18 examinations no lymph node was found in the region of highest activity. Fh-SPECT reconstruction artifacts occurred in 6/18 examinations.

CONCLUSION:

The fusion imaging concept combining SLN information with ultrasound images presented here proves to be feasible and technically successful. However, significant technical limitations were shown in fh-SPECT quality and fusion precision. Subject to technical optimisation of SPECT quality and co-registration, a meaningful contribution to the preoperative planning of lymph node therapy is imaginable. Thus, fundamentally a preoperative histological examination by fh-SPECT/US-guided biopsy is possible.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Mama / Tomografia Computadorizada de Emissão de Fóton Único / Ultrassonografia / Linfonodos / Melanoma Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Nuklearmedizin Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Mama / Tomografia Computadorizada de Emissão de Fóton Único / Ultrassonografia / Linfonodos / Melanoma Tipo de estudo: Diagnostic_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Nuklearmedizin Ano de publicação: 2014 Tipo de documento: Article