Performance and System Validation of a New Cellular-Enabled Blood Glucose Monitoring System Using a New Standard Reference Measurement Procedure of Isotope Dilution UPLC-MRM Mass Spectrometry.
Angelides, Kimon; Matsunami, Risë K; Engler, David A.
Afiliação
  • Angelides K; Livongo Health, Inc, Palo Alto, CA, USA kangelides@livongo.com.
  • Matsunami RK; Proteomics Programmatic Core Laboratory, Houston Methodist Hospital Research Institute, Houston, TX, USA.
  • Engler DA; Proteomics Programmatic Core Laboratory, Houston Methodist Hospital Research Institute, Houston, TX, USA.
J Diabetes Sci Technol ; 9(5): 1061-70, 2015 May 22.
Article em En | MEDLINE | ID: mdl-26002836
ABSTRACT

BACKGROUND:

We evaluated the accuracy, precision, and linearity of the In Touch blood glucose monitoring system (BGMS), a new color touch screen and cellular-enabled blood glucose meter, using a new rapid, highly precise and accurate (13)C6 isotope-dilution liquid chromatography-mass spectrometry method (IDLC-MS).

METHODS:

Blood glucose measurements from the In Touch BGMS were referenced to a validated UPLC-MRM standard reference measurement procedure previously shown to be highly accurate and precise. Readings from the In Touch BGMS were taken over the blood glucose range of 24-640 mg/dL using 12 concentrations of blood glucose. Ten In Touch BGMS and 3 lots of test strips were used with 10 replicates at each concentration. A lay user study was also performed to assess the ease of use.

RESULTS:

At blood glucose concentrations <75 mg/dL 100% of the measurements are within ±8 mg/dL from the true reference standard; at blood glucose levels >75 mg/dL 100% of the measurements are within ±15% of the true reference standard. 100% of the results are within category A of the consensus grid. Within-run precision show CV < 3.72% between 24-50 mg/dL and CV<2.22% between 500 and 600 mg/dL. The results show that the In Touch meter exceeds the minimum criteria of both the ISO 151972003 and ISO 151972013 standards. The results from a user panel show that 100% of the respondents reported that the color touch screen, with its graphic user interface (GUI), is well labeled and easy to navigate.

CONCLUSIONS:

To our knowledge this is the first touch screen glucose meter and the first study where accuracy of a new BGMS has been measured against a true primary reference standard, namely IDLC-MS.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glicemia / Automonitorização da Glicemia / Diabetes Mellitus Tipo 1 / Diabetes Mellitus Tipo 2 Limite: Humans Idioma: En Revista: J Diabetes Sci Technol Assunto da revista: ENDOCRINOLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glicemia / Automonitorização da Glicemia / Diabetes Mellitus Tipo 1 / Diabetes Mellitus Tipo 2 Limite: Humans Idioma: En Revista: J Diabetes Sci Technol Assunto da revista: ENDOCRINOLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Estados Unidos