Your browser doesn't support javascript.
loading
Minimizing the effects of magnetization transfer asymmetry on inhomogeneous magnetization transfer (ihMT) at ultra-high magnetic field (11.75 T).
Prevost, Valentin H; Girard, Olivier M; Varma, Gopal; Alsop, David C; Duhamel, Guillaume.
Afiliação
  • Prevost VH; Centre de Résonance Magnétique Biologique et Médicale, CRMBM-CEMEREM, UMR 7339, CNRS, Faculté de Médecine, Aix-Marseille Université (AMU), 27 Boulevard Jean Moulin, 13005, Marseille, France.
  • Girard OM; Centre de Résonance Magnétique Biologique et Médicale, CRMBM-CEMEREM, UMR 7339, CNRS, Faculté de Médecine, Aix-Marseille Université (AMU), 27 Boulevard Jean Moulin, 13005, Marseille, France.
  • Varma G; Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, USA.
  • Alsop DC; Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, USA.
  • Duhamel G; Centre de Résonance Magnétique Biologique et Médicale, CRMBM-CEMEREM, UMR 7339, CNRS, Faculté de Médecine, Aix-Marseille Université (AMU), 27 Boulevard Jean Moulin, 13005, Marseille, France. guillaume.duhamel@univ-amu.fr.
MAGMA ; 29(4): 699-709, 2016 Aug.
Article em En | MEDLINE | ID: mdl-26762244
ABSTRACT

OBJECTIVES:

The recently reported inhomogeneous magnetization transfer technique (ihMT) has been proposed for specific imaging of inhomogeneously broadened lines, and has shown great promise for characterizing myelinated tissues. The ihMT contrast is obtained by subtracting magnetization transfer images obtained with simultaneous saturation at positive and negative frequency offsets (dual frequency saturation experiment, MT (+/-)) from those obtained with single frequency saturation (MT (+)) at the same total power. Hence, ihMT may be biased by MT-asymmetry, especially at ultra-high magnetic field. Use of the average of single positive and negative frequency offset saturation MT images, i.e., (MT (+)+MT (-)) has been proposed to correct the ihMT signal from MT-asymmetry signal. MATERIALS AND

METHODS:

The efficiency of this correction method was experimentally assessed in this study, performed at 11.75 T on mice. Quantitative corrected ihMT and MT-asymmetry ratios (ihMTR and MTRasym) were measured in mouse brain structures for several MT-asymmetry magnitudes and different saturation parameter sets.

RESULTS:

Our results indicated a "safe" range of magnitudes (/MTRasym/<4 %) for which MT-asymmetry signal did not bias the corrected ihMT signal. Moreover, experimental evidence of the different natures of both MT-asymmetry and inhomogeneous MT contrasts were provided. In particular, non-zero ihMT ratios were obtained at zero MTRasym values.

CONCLUSION:

MTRasym is not a confounding factor for ihMT quantification, even at ultra-high field, as long as MTRasym is restricted to ±4 %.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Campos Magnéticos / Magnetismo / Bainha de Mielina Limite: Animals Idioma: En Revista: MAGMA Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: França

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Campos Magnéticos / Magnetismo / Bainha de Mielina Limite: Animals Idioma: En Revista: MAGMA Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: França