Your browser doesn't support javascript.
loading
Dual-pathway multi-echo sequence for simultaneous frequency and T2 mapping.
Cheng, Cheng-Chieh; Mei, Chang-Sheng; Duryea, Jeffrey; Chung, Hsiao-Wen; Chao, Tzu-Cheng; Panych, Lawrence P; Madore, Bruno.
Afiliação
  • Cheng CC; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, USA.
  • Mei CS; Department of Physics, Soochow University, Taipei, Taiwan.
  • Duryea J; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, USA.
  • Chung HW; Graduate Institute of Biomedical Electronics and Bioinformatics, National Taiwan University, Taipei, Taiwan.
  • Chao TC; Department of Computer Science and Information Engineering, National Cheng-Kung University, Tainan, Taiwan.
  • Panych LP; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, USA.
  • Madore B; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, USA.
J Magn Reson ; 265: 177-87, 2016 Apr.
Article em En | MEDLINE | ID: mdl-26923150
ABSTRACT

PURPOSE:

To present a dual-pathway multi-echo steady state sequence and reconstruction algorithm to capture T2, T2(∗) and field map information.

METHODS:

Typically, pulse sequences based on spin echoes are needed for T2 mapping while gradient echoes are needed for field mapping, making it difficult to jointly acquire both types of information. A dual-pathway multi-echo pulse sequence is employed here to generate T2 and field maps from the same acquired data. The approach might be used, for example, to obtain both thermometry and tissue damage information during thermal therapies, or susceptibility and T2 information from a same head scan, or to generate bonus T2 maps during a knee scan.

RESULTS:

Quantitative T2, T2(∗) and field maps were generated in gel phantoms, ex vivo bovine muscle, and twelve volunteers. T2 results were validated against a spin-echo reference standard A linear regression based on ROI analysis in phantoms provided close agreement (slope/R(2)=0.99/0.998). A pixel-wise in vivo Bland-Altman analysis of R2=1/T2 showed a bias of 0.034 Hz (about 0.3%), as averaged over four volunteers. Ex vivo results, with and without motion, suggested that tissue damage detection based on T2 rather than temperature-dose measurements might prove more robust to motion.

CONCLUSION:

T2, T2(∗) and field maps were obtained simultaneously, from the same datasets, in thermometry, susceptibility-weighted imaging and knee-imaging contexts.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador Tipo de estudo: Diagnostic_studies Limite: Animals / Humans Idioma: En Revista: J Magn Reson Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador Tipo de estudo: Diagnostic_studies Limite: Animals / Humans Idioma: En Revista: J Magn Reson Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos