Premature Trigger of ERI in Medtronic EnRhythm Devices.
Middeldorp, Melissa E; Mahajan, Rajiv; Elliott, Adrian D; Pathak, Rajeev K; Twomey, Darragh; Wilson, Lauren; Stolcman, Simon; Munawar, Dian A; Kumar, Sharath; Lau, Dennis H; Sanders, Prashanthan.
Afiliação
  • Middeldorp ME; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Mahajan R; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Elliott AD; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Pathak RK; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Twomey D; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Wilson L; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Stolcman S; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Munawar DA; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Kumar S; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Lau DH; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
  • Sanders P; Centre for Heart Rhythm Disorders (CHRD), South Australian Health and Medical Research Institute (SAHMRI), University of Adelaide and Royal Adelaide Hospital, Adelaide, Australia.
Pacing Clin Electrophysiol ; 40(6): 624-628, 2017 Jun.
Article em En | MEDLINE | ID: mdl-28294359
ABSTRACT

BACKGROUND:

Medical technology has made significant advances over the last few decades with smaller and more dynamic pacemakers. However, technical failures leading to premature replacement is a cause of concern. We present a series of Medtronic EnRhythm devices that reached premature elective replacement indicator (ERI).

METHODS:

The database of Centre of Heart Rhythm Disorders was searched for EnRhythm device implantation from 2006 to 2011. Battery depletion <8.5 years was considered premature considering the projected average longevity to be 8.5-10.5 years. An unexpected premature ERI was defined when it was reached within 3 months of last normal check. Device follow-up was conducted every 3 months after advisory.

RESULTS:

A total of 88 EnRhythm pacemakers were implanted. Over a median follow-up of 6.2 years (range 0.3-9.2), 39 (44.3%) EnRhythm devices reached premature ERI. In 11 (28%), ERI was not recognized and patients were being investigated for other causes of unsteadiness or dyspnea prior to device check. Notably, three (7%) patients had premature ERI < 3.5 years. Ten (25.6%) had sudden and unexpected premature ERI. While asynchronous pacing was observed, there were no cases of absence of pacing.

CONCLUSIONS:

The rate of premature ERI for EnRhythm devices was 44.3%, significantly higher than reported by the manufacturer. Of concern, a sizeable proportion occurred unexpectedly, warranting more frequent reviews and empirical replacement in some patients. With the experience of the EnRhythm, appropriate monitoring strategies are recommended for future advisories.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Fontes de Energia Elétrica / Análise de Falha de Equipamento / Remoção de Dispositivo / Falha de Equipamento / Segurança de Equipamentos / Recall de Dispositivo Médico Limite: Aged / Female / Humans / Male País/Região como assunto: Oceania Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Austrália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Fontes de Energia Elétrica / Análise de Falha de Equipamento / Remoção de Dispositivo / Falha de Equipamento / Segurança de Equipamentos / Recall de Dispositivo Médico Limite: Aged / Female / Humans / Male País/Região como assunto: Oceania Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Austrália