Does the Additional Articulation in Retrieved Bipolar Hemiarthroplasty Implants Decrease Trunnionosis Compared to Total Hip Arthroplasty?
Del Balso, Christopher; Teeter, Matthew G; Tan, Sok C; Lanting, Brent A; Howard, James L.
Afiliação
  • Del Balso C; Division of Orthopaedic Surgery, Department of Surgery, Schulich School of Medicine and Dentistry, Western University, London Health Sciences Centre, University Hospital, London, Ontario, Canada.
  • Teeter MG; Department of Medical Biophysics, Schulich School of Medicine & Dentistry, Western University, London, Ontario, Canada; Surgical Innovation Program, Lawson Health Research Institute, London, Ontario, Canada.
  • Tan SC; Division of Orthopaedic Surgery, Department of Surgery, Schulich School of Medicine and Dentistry, Western University, London Health Sciences Centre, University Hospital, London, Ontario, Canada.
  • Lanting BA; Division of Orthopaedic Surgery, Department of Surgery, Schulich School of Medicine and Dentistry, Western University, London Health Sciences Centre, University Hospital, London, Ontario, Canada.
  • Howard JL; Division of Orthopaedic Surgery, Department of Surgery, Schulich School of Medicine and Dentistry, Western University, London Health Sciences Centre, University Hospital, London, Ontario, Canada.
J Arthroplasty ; 33(1): 268-272, 2018 01.
Article em En | MEDLINE | ID: mdl-29033155
ABSTRACT

BACKGROUND:

Trunnionosis at the modular head-neck taper interface in metal-on-polyethylene total hip arthroplasty (MoP THA) has been shown to occur, and represents a potential mode of MoP THA failure. The purpose of the present investigation is to elucidate differences in fretting and corrosion at the head-neck taper interface of prostheses retrieved from bipolar hemiarthroplasty (BH) and MoP THA.

METHODS:

A retrieval analysis of BH and MoP THA prostheses featuring a single taper design from a single manufacturer and in vivo for a minimum 2 years was performed. Fifteen femoral heads of 28-mm diameter and corresponding femoral stems retrieved from BH were compared with MoP THA implants matched based on time in vivo and head length (28 mm, -3 mm to 28 mm, +8 mm). Fretting and corrosion damage scoring was completed under stereomicroscopic visualization.

RESULTS:

Femoral head bore tapers retrieved from BH exhibited decreased overall fretting (P = .02), when compared to those retrieved from MoP THA. Total corrosion scores for all retrieved implants were positively correlated with implantation time (ρ = 0.54, P < .02).

CONCLUSION:

Femoral heads retrieved from BH exhibit decreased fretting damage compared to those retrieved from MoP THA. The added articulation in BH implants may decrease torque produced at the head-neck taper junction, thereby decreasing fretting. Increased fretting damage in implants from MoP THA is not associated with increased corrosion in 28-mm heads of this taper design. The longer a BH or MoP THA prosthesis is implanted, the greater the risk of damage due to corrosion.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Falha de Prótese / Artroplastia de Quadril / Hemiartroplastia / Prótese de Quadril Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Falha de Prótese / Artroplastia de Quadril / Hemiartroplastia / Prótese de Quadril Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Canadá