Your browser doesn't support javascript.
loading
In-vivo postoperative motion analysis of metal and ceramic bipolar hip hemiarthroplasty.
Yoshioka, Toru; Okimoto, Nobukazu; Fuse, Yoshifumi; Kawasaki, Makoto; Mori, Toshiharu; Sakai, Akinori; Majima, Tokifumi.
Afiliação
  • Yoshioka T; Department of Orthopaedic Surgery, Shimura Hospital, 3-13 Funairimachi, Naka-ku, Hiroshima-shi, Hiroshima, 731-0841, Japan. Electronic address: t-yossy@rc4.so-net.ne.jp.
  • Okimoto N; Okimoto Clinic, 185-4 Yutakamachi Kubi, Kure-shi, Hiroshima, 734-0304, Japan.
  • Fuse Y; Department of Orthopaedic Surgery, Sakamidorii Hospital, 6-28-1 Midorii, Asaminami-ku, Hiroshima-shi, Hiroshima, 731-0103, Japan.
  • Kawasaki M; Department of Orthopaedic Surgery, University of Occupational and Environmental Health, 1-1 Iseigaoka, Yahatanishi-ku, Kitakyushu-shi, Fukuoka, 807-0804, Japan.
  • Mori T; Department of Orthopaedic Surgery, Shinkokura Hospital, 1-3-1 Kaneda, Kokurakita-ku, Kitakyushu-shi, Fukuoka, 803-8505, Japan.
  • Sakai A; Department of Orthopaedic Surgery, University of Occupational and Environmental Health, 1-1 Iseigaoka, Yahatanishi-ku, Kitakyushu-shi, Fukuoka, 807-0804, Japan.
  • Majima T; Department of Orthopaedic Surgery, Nippon Medical School, 1-1-5 Sendagi, Bunkyo-ku, Tokyo, 113-8602, Japan.
J Orthop Sci ; 23(2): 371-376, 2018 Mar.
Article em En | MEDLINE | ID: mdl-29229239
ABSTRACT

BACKGROUND:

We hypothesized that ceramic bipolar shows less impingement between stem neck and outer head compared to metal bipolar, because of its low coefficient of friction with the acetabulum cartilage. In this paper, a three dimensional (3D) postoperative motion analysis has been used to assess the different motion characteristics of metal and ceramic bipolar hip hemiarthroplasty systems.

METHODS:

This study was conducted on 40 patients divided in two matched cohorts 20 patients with metal bipolar and 20 patients with the ceramic bipolar. We obtained motion pictures from standing position to maximum abduction in flexion by fluoroscopy then analyzed by 2D-3D image matching method. The motion range of the "Outer head angle", "Stem neck angle" and the "Stem neck and outer head angle" was compared between the metal bipolar group and the ceramic bipolar group.

RESULTS:

The metal bipolar group's inner head's range of movement was greater than the ceramic bipolar group. Impingement between stem neck and outer head occurred in 30% of metal group patients. There were no impingement cases for the ceramic bipolar group.

CONCLUSIONS:

The ceramic bipolar shows less impingement between stem neck and outer head compared to the metal bipolar. Ceramic bipolar may reduce the typical bipolar related complication and exert less effect on the acetabular cartilage due to less surface coefficient of friction.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Amplitude de Movimento Articular / Imageamento Tridimensional / Hemiartroplastia / Articulação do Quadril / Prótese de Quadril Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Orthop Sci Assunto da revista: ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Amplitude de Movimento Articular / Imageamento Tridimensional / Hemiartroplastia / Articulação do Quadril / Prótese de Quadril Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Orthop Sci Assunto da revista: ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article