Biomechanical evaluation of an arthroscopic transosseous repair as a revision option for failed rotator cuff surgery.
Dyrna, Felix; Voss, Andreas; Pauzenberger, Leo; Obopilwe, Elifho; Mazzocca, Augustus D; Castagna, Alessandro; Edgar, Cory.
Afiliação
  • Dyrna F; Department of Orthopaedic Sports Medicine, Technical University, Munich, Germany.
  • Voss A; Department of Orthopaedic Sports Medicine, Technical University, Munich, Germany.
  • Pauzenberger L; Department of Orthopaedic Surgery, University of Connecticut, Farmington, CT, USA.
  • Obopilwe E; Department of Orthopaedic Surgery, University of Connecticut, Farmington, CT, USA.
  • Mazzocca AD; Department of Orthopaedic Surgery, University of Connecticut, Farmington, CT, USA.
  • Castagna A; Department of Shoulder and Elbow IRCCS Humanitas Institute, Milan, Italy.
  • Edgar C; Department of Orthopaedic Surgery, University of Connecticut, Farmington, CT, USA. coedgar@uchc.edu.
BMC Musculoskelet Disord ; 19(1): 240, 2018 Jul 19.
Article em En | MEDLINE | ID: mdl-30025529
ABSTRACT

BACKGROUND:

The number of revision rotator cuff cases is increasing. The literature is lacking guidance or biomechanical evaluation for fixation strength in a revision case scenario. Therefore, the aim of the study was to provide biomechanical data investigating primary fixation strength of a transosseous technique after anchor pullout failure of a single row reconstruction. It was hypothesized that an arthroscopic transosseous repair system as a procedure for rotator cuff revisions is providing equivalent stability compared to a primary single row suture anchor fixation due to change of fixation site.

METHODS:

Eight matched pairs (n = 16) of fresh frozen human shoulders were tested. The paired specimen shoulders were randomly divided into two repair groups (A single row and B primary transosseous repair). The potted specimens were mounted onto the Servohydraulic test system. Both groups were tested under cyclic loading followed by load to failure testing. Suture anchor repair shoulders (group A) that were tested to failure underwent a revision transosseous repair and were subsequently tested again using the same setup and protocol (group C).

RESULTS:

The mean native footprint areas did not show a significant difference between groups. The reconstructed footprint area showed a significantly greater coverage in favor of the transosseous repair. Ultimate load to failure of reconstructions with the primary anchor fixation (344.73 N ± 63.19) and the primary transosseous device (375.36 N ± 70.27) was not significantly higher compared to the revision repair (332.19 N ± 119.01 p = 0.45, p = 0.53).

CONCLUSION:

The tested transosseous anchor device is a suitable option to widely used suture anchors, providing equivalent fixation properties even in a revision case scenario. LEVEL OF EVIDENCE Basic Science Study, Biomechanics.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroscopia / Reoperação / Articulação do Ombro / Manguito Rotador Tipo de estudo: Evaluation_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Artroscopia / Reoperação / Articulação do Ombro / Manguito Rotador Tipo de estudo: Evaluation_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Alemanha