Quantification of Eccentric Fixation Using Spectral-Domain Optical Coherence Tomography.
Nakamoto, Yuki; Takada, Ryota; Tanaka, Minatsu; Matsumoto, Fumiko; Aomatsu, Keiichi; Tachibana, Kuniko; Eguchi, Hiroshi; Kusaka, Shunji.
Afiliação
  • Nakamoto Y; Department of Ophthalmology, Kindai University Sakai Hospital, Sakai, Japan.
  • Takada R; Department of Ophthalmology, Kindai University Sakai Hospital, Sakai, Japan.
  • Tanaka M; Department of Ophthalmology, Kindai University Sakai Hospital, Sakai, Japan.
  • Matsumoto F; Department of Ophthalmology, Kindai University Sakai Hospital, Sakai, Japan.
  • Aomatsu K; Department of Ophthalmology, Kindai University Hospital, Osakasayama, Japan.
  • Tachibana K; Department of Ophthalmology, Kindai University Sakai Hospital, Sakai, Japan.
  • Eguchi H; Department of Ophthalmology, Kindai University Sakai Hospital, Sakai, Japanhiroegu0113@gmail.com.
  • Kusaka S; Department of Ophthalmology, Kindai University Sakai Hospital, Sakai, Japan.
Ophthalmic Res ; 60(4): 231-237, 2018.
Article em En | MEDLINE | ID: mdl-30326478
ABSTRACT

PURPOSE:

This study aimed to evaluate the utility of optical coherence tomography (OCT) in the quantification of eccentric fixation in amblyopic patients. MATERIAL AND

METHODS:

In this study, 14 amblyopic patients and 10 healthy volunteers were enrolled. Under non-mydriatic conditions, fixation tests were performed directly using a fixation ophthalmoscope and indirectly using spectral-domain OCT. For evaluations using OCT, the distance between the fovea and the fixation point, which was determined by a cross-sectional image, was measured.

RESULTS:

On evaluations of healthy volunteers by OCT, the mean distance between the fixation point and the fovea was 80.4 ± 37.7 µm for the dominant eyes and 63.7 ± 36.4 µm for non-dominant eyes (p = 0.41). In amblyopic patients, on evaluation by OCT, the mean distance between the fixation point and the fovea was 193.8 ± 188.3 µm in amblyopic eyes and 83.5 ± 39.3 µm in paired fellow eyes (p = 0.02). Although OCT could detect eccentric fixation points in all the affected eyes of amblyopic patients, fixation ophthalmoscope was unable to quantify them in 2 of 14 affected eyes.

CONCLUSIONS:

Compared with a fixation ophthalmoscope, our method using OCT seems to be superior both in quantification and detection of eccentric fixation in amblyopic patients, without the need for mydriasis.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Acuidade Visual / Ambliopia / Tomografia de Coerência Óptica / Fixação Ocular / Fóvea Central Tipo de estudo: Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Male Idioma: En Revista: Ophthalmic Res Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Acuidade Visual / Ambliopia / Tomografia de Coerência Óptica / Fixação Ocular / Fóvea Central Tipo de estudo: Observational_studies / Prevalence_studies / Risk_factors_studies Limite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Male Idioma: En Revista: Ophthalmic Res Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Japão