Your browser doesn't support javascript.
loading
Varus or valgus positioning of the tibial component of a unicompartmental fixed-bearing knee arthroplasty does not increase wear.
Woiczinski, Matthias; Schröder, Christian; Paulus, Alexander; Kistler, Manuel; Jansson, Volkmar; Müller, Peter E; Weber, Patrick.
Afiliação
  • Woiczinski M; Department of Orthopedic Surgery, Physical Medicine and Rehabilitation, University Hospital, Ludwig-Maximilians-University (LMU), Campus Großhadern, Marchioninistr. 15, 81377, Munich, Germany.
  • Schröder C; Department of Orthopedic Surgery, Physical Medicine and Rehabilitation, University Hospital, Ludwig-Maximilians-University (LMU), Campus Großhadern, Marchioninistr. 15, 81377, Munich, Germany.
  • Paulus A; Department of Orthopedic Surgery, Physical Medicine and Rehabilitation, University Hospital, Ludwig-Maximilians-University (LMU), Campus Großhadern, Marchioninistr. 15, 81377, Munich, Germany.
  • Kistler M; Department of Orthopedic Surgery, Physical Medicine and Rehabilitation, University Hospital, Ludwig-Maximilians-University (LMU), Campus Großhadern, Marchioninistr. 15, 81377, Munich, Germany.
  • Jansson V; Department of Orthopedic Surgery, Physical Medicine and Rehabilitation, University Hospital, Ludwig-Maximilians-University (LMU), Campus Großhadern, Marchioninistr. 15, 81377, Munich, Germany.
  • Müller PE; Department of Orthopedic Surgery, Physical Medicine and Rehabilitation, University Hospital, Ludwig-Maximilians-University (LMU), Campus Großhadern, Marchioninistr. 15, 81377, Munich, Germany.
  • Weber P; Department of Orthopedic Surgery, Physical Medicine and Rehabilitation, University Hospital, Ludwig-Maximilians-University (LMU), Campus Großhadern, Marchioninistr. 15, 81377, Munich, Germany. patrick.weber@med.uni-muenchen.de.
Knee Surg Sports Traumatol Arthrosc ; 28(9): 3016-3021, 2020 Sep.
Article em En | MEDLINE | ID: mdl-31690992
ABSTRACT

PURPOSE:

Higher revision rates were shown in varus- or valgus-positioned tibias in unicompartmental knee arthroplasty (UKA), but more than 15% of UKA prostheses are implanted with more than 5° of varus or valgus. This study aimed to analyze the wear rate in UKA when implanting the tibial component in either varus or valgus position versus a neutral placement at 90° to the tibial anatomical axis. The study hypothesized that a 5° varus or valgus positioning of the tibial plateau will generate less wear compared to a neutral alignment.

METHODS:

Wear was experimentally analyzed on a medial anatomical fixed-bearing unicompartmental knee prosthesis (Univation, Aesculap, Germany) in vitro with a customized, four-station, servohydraulic knee wear simulator, reproducing the walking cycle. The forces, loading and range of motion were applied as specified in the ISO 14243-12002, 5 million cycles were analyzed. The tibial components of the medial prostheses were inserted in a neutral position, with 5° varus, and 5° valgus (n = 3, each group).

RESULTS:

The wear rate decreased significantly with a 5° varus positioning (6.30 ± 1.38 mg/million cycles) and a 5° valgus positioning (4.96 ± 2.47 mg/million cycles) compared to the neutral position (12.16 ± 1.26 mg/million cycles) (p < 0.01 for the varus and the valgus position). The wear area on the inlay was slightly reduced in the varus and valgus group.

CONCLUSION:

A varus or valgus "malpositioning" up to 5° will not lead to an increased wear. Wear was even less because of the reduced articulating contact area between the inlay and the femur. A slight varus positioning of the tibial component (parallel to the anatomical joint line) positioning can be advocated from a point of wear. LEVEL OF EVIDENCE Experimental study.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Teste de Materiais / Falha de Prótese / Ajuste de Prótese / Artroplastia do Joelho / Prótese do Joelho Limite: Humans Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Assunto da revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Teste de Materiais / Falha de Prótese / Ajuste de Prótese / Artroplastia do Joelho / Prótese do Joelho Limite: Humans Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Assunto da revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Alemanha