Twin Robotic X-Ray System for 3D Cone-Beam CT of the Wrist: An Evaluation of Image Quality and Radiation Dose.
Grunz, Jan-Peter; Gietzen, Carsten H; Kunz, Andreas S; Weng, Andreas M; Veyhl-Wichmann, Maike; Ergün, Süleyman; Bley, Thorsten A; Schmitt, Rainer; Gassenmaier, Tobias.
Afiliação
  • Grunz JP; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacherstr 6, 97080 Würzburg, Germany.
  • Gietzen CH; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacherstr 6, 97080 Würzburg, Germany.
  • Kunz AS; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacherstr 6, 97080 Würzburg, Germany.
  • Weng AM; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacherstr 6, 97080 Würzburg, Germany.
  • Veyhl-Wichmann M; Institute of Anatomy and Cell Biology, University of Würzburg, Würzburg, Germany.
  • Ergün S; Institute of Anatomy and Cell Biology, University of Würzburg, Würzburg, Germany.
  • Bley TA; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacherstr 6, 97080 Würzburg, Germany.
  • Schmitt R; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacherstr 6, 97080 Würzburg, Germany.
  • Gassenmaier T; Department of Diagnostic and Interventional Radiology, Rhön-Klinikum Campus Bad Neustadt, Bad Neustadt an der Saale, Germany.
AJR Am J Roentgenol ; 214(2): 422-427, 2020 02.
Article em En | MEDLINE | ID: mdl-31799871
ABSTRACT
OBJECTIVE. The purpose of this study was to assess image quality and radiation dose of a novel twin robotic x-ray system's 3D cone-beam CT (CBCT) function for the depiction of cadaveric wrists. MATERIALS AND METHODS. Sixteen cadaveric wrists were scanned using dedicated low-dose and standard-dose CBCT protocols as well as clinical MDCT for comparison. Three readers assessed overall image quality, noise, and artifacts in bone and soft tissue on 5-point Likert scales. For radiation dose analysis, volume CT dose indexes (CTDIvol) were compared. RESULTS. Overall image quality of most studies was very good or excellent in MDCT (for readers 1, 2, and 3 100%, 100%, and 88%, respectively), standard-dose CBCT (100%, 100%, and 94%), and low dose CBCT (100%, 94%, and 88%) with two readers favoring standard-dose CBCT over MDCT image quality (readers 1 and 2; p ≤ 0.046). In soft tissue, standard-dose (readers 1, 2, and 3; p ≤ 0.021) and low-dose (all p ≤ 0.001) CBCT images had more noise than MDCT in all cases. Standard-dose (all p ≤ 0.003) and low-dose (all p < 0.001) CBCT images also displayed more artifacts. In osseous tissue, one reader observed more noise (p < 0.001) and artifacts (p = 0.020) for low-dose CBCT than for MDCT, whereas no difference was found between standard-dose CBCT and MDCT. Mean CTDIvol was significantly lower for standard-dose (5.2 ± 0.6 mGy; p < 0.001) and low-dose CBCT (1.8 ± 0.2 mGy; p < 0.001) than for clinical MDCT without automatic dose modulation (15.0 ± 0.0 mGy). CONCLUSION. The tested CBCT function delivers suitable image quality for clinical wrist imaging at significantly lower radiation levels than conventional MDCT. In combination with comfortable positioning options and the ability to perform additional radiographic and fluoroscopic examinations, the twin robotic x-ray system may hold the potential to be a one-stop shop device for trauma-associated wrist imaging.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Articulação do Punho / Robótica / Imageamento Tridimensional / Tomografia Computadorizada de Feixe Cônico Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: AJR Am J Roentgenol Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doses de Radiação / Articulação do Punho / Robótica / Imageamento Tridimensional / Tomografia Computadorizada de Feixe Cônico Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: AJR Am J Roentgenol Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Alemanha