Design and testing of a 96-channel neural interface module for the Networked Neuroprosthesis system.
Bullard, Autumn J; Nason, Samuel R; Irwin, Zachary T; Nu, Chrono S; Smith, Brian; Campean, Alex; Peckham, P Hunter; Kilgore, Kevin L; Willsey, Matthew S; Patil, Parag G; Chestek, Cynthia A.
Afiliação
  • Bullard AJ; 1Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI USA.
  • Nason SR; 1Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI USA.
  • Irwin ZT; 1Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI USA.
  • Nu CS; 1Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI USA.
  • Smith B; 2Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH USA.
  • Campean A; 2Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH USA.
  • Peckham PH; 2Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH USA.
  • Kilgore KL; 3Department of Orthopaedics, MetroHealth Medical Center, Cleveland, OH USA.
  • Willsey MS; 2Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH USA.
  • Patil PG; 3Department of Orthopaedics, MetroHealth Medical Center, Cleveland, OH USA.
  • Chestek CA; 4Research Service, Louis Stokes Cleveland Department of Veterans Affairs Medical Center, Cleveland, OH USA.
Bioelectron Med ; 5: 3, 2019.
Article em En | MEDLINE | ID: mdl-32232094
ABSTRACT

BACKGROUND:

The loss of motor functions resulting from spinal cord injury can have devastating implications on the quality of one's life. Functional electrical stimulation has been used to help restore mobility, however, current functional electrical stimulation (FES) systems require residual movements to control stimulation patterns, which may be unintuitive and not useful for individuals with higher level cervical injuries. Brain machine interfaces (BMI) offer a promising approach for controlling such systems; however, they currently still require transcutaneous leads connecting indwelling electrodes to external recording devices. While several wireless BMI systems have been designed, high signal bandwidth requirements limit clinical translation. Case Western Reserve University has developed an implantable, modular FES system, the Networked Neuroprosthesis (NNP), to perform combinations of myoelectric recording and neural stimulation for controlling motor functions. However, currently the existing module capabilities are not sufficient for intracortical recordings.

METHODS:

Here we designed and tested a 1 × 4 cm, 96-channel neural recording module prototype to fit within the specifications to mate with the NNP. The neural recording module extracts power between 0.3-1 kHz, instead of transmitting the raw, high bandwidth neural data to decrease power requirements.

RESULTS:

The module consumed 33.6 mW while sampling 96 channels at approximately 2 kSps. We also investigated the relationship between average spiking band power and neural spike rate, which produced a maximum correlation of R = 0.8656 (Monkey N) and R = 0.8023 (Monkey W).

CONCLUSION:

Our experimental results show that we can record and transmit 96 channels at 2ksps within the power restrictions of the NNP system and successfully communicate over the NNP network. We believe this device can be used as an extension to the NNP to produce a clinically viable, fully implantable, intracortically-controlled FES system and advance the field of bioelectronic medicine.