Fast quantitative three-dimensional ultrashort echo time (UTE) Cones magnetic resonance imaging of major tissues in the knee joint using extended sprial sampling.
Wan, Lidi; Ma, Yajun; Yang, Jiawei; Jerban, Saeed; Searleman, Adam C; Carl, Michael; Le, Nicole; Chang, Eric Y; Tang, Guangyu; Du, Jiang.
Afiliação
  • Wan L; Department of Radiology, Tenth People's Hospital of Tongji University, Shanghai, China.
  • Ma Y; Department of Radiology, University of California, San Diego, California.
  • Yang J; Department of Radiology, University of California, San Diego, California.
  • Jerban S; Department of Radiology, Tenth People's Hospital of Tongji University, Shanghai, China.
  • Searleman AC; Department of Radiology, University of California, San Diego, California.
  • Carl M; Department of Radiology, University of California, San Diego, California.
  • Le N; GE Healthcare, San Diego, California.
  • Chang EY; Radiology Service, Veterans Affairs San Diego Healthcare System, San Diego, California.
  • Tang G; Department of Radiology, University of California, San Diego, California.
  • Du J; Radiology Service, Veterans Affairs San Diego Healthcare System, San Diego, California.
NMR Biomed ; 33(10): e4376, 2020 10.
Article em En | MEDLINE | ID: mdl-32667115
ABSTRACT
The purpose of this study is to investigate the effect of extending the spiral sampling window on quantitative 3D ultrashort echo time (UTE) Cones imaging of major knee joint tissues including articular cartilage, menisci, tendons and ligaments at 3 T. Nine cadaveric human whole knee specimens were imaged on a 3 T clinical MRI scanner. A series of quantitative 3D UTE Cones imaging biomarkers including T2 *, T1 , adiabatic T1ρ , magnetization transfer ratio (MTR) and macromolecular fraction (MMF) were estimated using spiral sampling trajectories with various durations. Errors in UTE MRI biomarkers as a function of sampling time were evaluated using a nonstretched spiral trajectory as a reference standard. No significant differences were observed by increasing the spiral sampling window from 1116 to 2232 µs in the calculated T2 *, T1 , adiabatic T1ρ , MTR and MMF, as all P-values were over .05 as assessed by ANOVA with two-sided Dunnett's test. Although extending the sampling window results in signal loss for short T2 components, there was limited effect on the calculated quantitative biomarkers, with error percentages typically smaller than 5% in all the evaluated tissues. The total scan time can be reduced by up to 54% with quantification errors of less than 5% in any evaluated major tissue in the knee joint, suggesting that 3D UTE Cones MRI techniques can be greatly accelerated by using a longer spiral sampling window without causing additional quantitative bias.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Imageamento Tridimensional / Articulação do Joelho Limite: Aged / Female / Humans / Male Idioma: En Revista: NMR Biomed Assunto da revista: DIAGNOSTICO POR IMAGEM / MEDICINA NUCLEAR Ano de publicação: 2020 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Imageamento Tridimensional / Articulação do Joelho Limite: Aged / Female / Humans / Male Idioma: En Revista: NMR Biomed Assunto da revista: DIAGNOSTICO POR IMAGEM / MEDICINA NUCLEAR Ano de publicação: 2020 Tipo de documento: Article País de afiliação: China