The importance of three dimensional coronary artery reconstruction accuracy when computing virtual fractional flow reserve from invasive angiography.
Solanki, Roshni; Gosling, Rebecca; Rammohan, Vignesh; Pederzani, Giulia; Garg, Pankaj; Heppenstall, James; Hose, D Rodney; Lawford, Patricia V; Narracott, Andrew J; Fenner, John; Gunn, Julian P; Morris, Paul D.
Afiliação
  • Solanki R; Department of Infection, Immunity and Cardiovascular Disease, The Medical School, University of Sheffield, Sheffield, UK.
  • Gosling R; Department of Infection, Immunity and Cardiovascular Disease, The Medical School, University of Sheffield, Sheffield, UK.
  • Rammohan V; Department of Cardiology, Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield, UK.
  • Pederzani G; Insigneo Institute for In Silico Medicine, University of Sheffield, Sheffield, UK.
  • Garg P; Department of Infection, Immunity and Cardiovascular Disease, The Medical School, University of Sheffield, Sheffield, UK.
  • Heppenstall J; Insigneo Institute for In Silico Medicine, University of Sheffield, Sheffield, UK.
  • Hose DR; Department of Infection, Immunity and Cardiovascular Disease, The Medical School, University of Sheffield, Sheffield, UK.
  • Lawford PV; Insigneo Institute for In Silico Medicine, University of Sheffield, Sheffield, UK.
  • Narracott AJ; Insigneo Institute for In Silico Medicine, University of Sheffield, Sheffield, UK.
  • Fenner J; Norwich Medical School, University of East Anglia, Norwich, UK.
  • Gunn JP; Department of Radiology, Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield, UK.
  • Morris PD; Department of Infection, Immunity and Cardiovascular Disease, The Medical School, University of Sheffield, Sheffield, UK.
Sci Rep ; 11(1): 19694, 2021 10 04.
Article em En | MEDLINE | ID: mdl-34608218
ABSTRACT
Three dimensional (3D) coronary anatomy, reconstructed from coronary angiography (CA), is now being used as the basis to compute 'virtual' fractional flow reserve (vFFR), and thereby guide treatment decisions in patients with coronary artery disease (CAD). Reconstruction accuracy is therefore important. Yet the methods required remain poorly validated. Furthermore, the magnitude of vFFR error arising from reconstruction is unkown. We aimed to validate a method for 3D CA reconstruction and determine the effect this had upon the accuracy of vFFR. Clinically realistic coronary phantom models were created comprosing seven standard stenoses in aluminium and 15 patient-based 3D-printed, imaged with CA, three times, according to standard clinical protocols, yielding 66 datasets. Each was reconstructed using epipolar line projection and intersection. All reconstructions were compared against the real phantom models in terms of minimal lumen diameter, centreline and surface similarity. 3D-printed reconstructions (n = 45) and the reference files from which they were printed underwent vFFR computation, and the results were compared. The average error in reconstructing minimum lumen diameter (MLD) was 0.05 (± 0.03 mm) which was < 1% (95% CI 0.13-1.61%) compared with caliper measurement. Overall surface similarity was excellent (Hausdorff distance 0.65 mm). Errors in 3D CA reconstruction accounted for an error in vFFR of ± 0.06 (Bland Altman 95% limits of agreement). Errors arising from the epipolar line projection method used to reconstruct 3D coronary anatomy from CA are small but contribute to clinically relevant errors when used to compute vFFR.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Angiografia Coronária / Vasos Coronários / Imageamento Tridimensional / Reserva Fracionada de Fluxo Miocárdico Tipo de estudo: Diagnostic_studies / Etiology_studies / Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: Sci Rep Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Angiografia Coronária / Vasos Coronários / Imageamento Tridimensional / Reserva Fracionada de Fluxo Miocárdico Tipo de estudo: Diagnostic_studies / Etiology_studies / Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: Sci Rep Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Reino Unido