Three-layer collagen-based vascular graft designed for low-flow peripheral vascular reconstructions.
Spacek, Miroslav; Chlup, Hynek; Mitas, Petr; Vesely, Jan; Lambert, Lukas; Mlcek, Mikulas; Krajicek, Milan; Lindner, Jaroslav; Grus, Tomas.
Afiliação
  • Spacek M; Charles University and General University Hospital in Prague, First Faculty of Medicine, 2nd Department of Surgery - Department of Cardiovascular Surgery, Prague, Czech Republic.
  • Chlup H; Czech Technical University in Prague, Faculty of Mechanical Engineering, Department of Mechanics, Biomechanics and Mechatronics, Prague, Czech Republic.
  • Mitas P; Charles University and General University Hospital in Prague, First Faculty of Medicine, 2nd Department of Surgery - Department of Cardiovascular Surgery, Prague, Czech Republic.
  • Vesely J; Czech Technical University in Prague, Faculty of Mechanical Engineering, Department of Mechanics, Biomechanics and Mechatronics, Prague, Czech Republic.
  • Lambert L; Charles University and General University Hospital in Prague, First Faculty of Medicine, Department of Radiology, Prague, Czech Republic.
  • Mlcek M; Charles University, First Faculty of Medicine, Institute of Physiology, Prague, Czech Republic.
  • Krajicek M; Charles University and General University Hospital in Prague, First Faculty of Medicine, 2nd Department of Surgery - Department of Cardiovascular Surgery, Prague, Czech Republic.
  • Lindner J; Charles University and General University Hospital in Prague, First Faculty of Medicine, 2nd Department of Surgery - Department of Cardiovascular Surgery, Prague, Czech Republic.
  • Grus T; Charles University and General University Hospital in Prague, First Faculty of Medicine, 2nd Department of Surgery - Department of Cardiovascular Surgery, Prague, Czech Republic.
J Appl Biomed ; 17(1): 52, 2019 Mar.
Article em En | MEDLINE | ID: mdl-34907746
ABSTRACT

INTRODUCTION:

The aim of this study was to develop a prototype of an artificial blood vessel which has similar mechanical properties to a human saphenous vein graft and to experimentally verify the function of the prosthesis via ovine carotid bypass implantation. MATERIAL AND

METHODS:

The prototype of an artificial graft prosthesis for low flow was developed and manufactured from a collagenous matrix and reinforcing polyester mesh. We compared the results of both the pressurisation and the mechanical stress evaluation tests of VSM with four types of hybrid vascular graft. The most similar graft (type II) was chosen for the first ovine model implantation.

RESULTS:

Dominant behavior e.g. mechanical response of VSM graft in plots of circumferential and axial stress during loading is observed in circumferential direction. Average results of used VSM showed area of ideal mechanical response and the properties of artificial blood vessels were fitted into this area. Developed graft remained patent after 161 days of follow up in ovine model.

CONCLUSIONS:

The mechanical properties of the graft were designed and adjusted to be similar to the behaviour of human saphenous veins. This approach showed promising results and enhanced the final performance of the prosthesis.