Your browser doesn't support javascript.
loading
Evaluation of a new method of calculating breast tumor volume based on automated breast ultrasound.
Ma, Jing-Jing; Meng, Shan; Dang, Sha-Jie; Wang, Jia-Zhong; Yuan, Quan; Yang, Qi; Song, Can-Xu.
Afiliação
  • Ma JJ; Department of Internal Medicine, Xi'an Fifth Hospital, Xi'an, China.
  • Meng S; Department of Hematology, The Second Affiliated Hospital of Xi'an Jiaotong University, Xi'an, China.
  • Dang SJ; Department of Anesthesia, Shaanxi Provincial Cancer Hospital, Affiliated to Xi'an Jiaotong University, Xi'an, China.
  • Wang JZ; Department of General Surgery, The Second Affiliated Hospital of Xi'an Jiaotong University, Xi'an, China.
  • Yuan Q; Department of Ultrasound, Shaanxi Provincial Cancer Hospital, Affiliated to Xi'an Jiaotong University, Xi'an, China.
  • Yang Q; Department of Surgery, Shaanxi Provincial Cancer Hospital, Affiliated to Xi'an Jiaotong University, Xi'an, China.
  • Song CX; Department of Ultrasound, Shaanxi Provincial Cancer Hospital, Affiliated to Xi'an Jiaotong University, Xi'an, China.
Front Oncol ; 12: 895575, 2022.
Article em En | MEDLINE | ID: mdl-36176389
ABSTRACT

Objective:

To evaluate the effectiveness and advantages of a new method for calculating breast tumor volume based on an automated breast ultrasound system (ABUS).

Methods:

A total of 42 patients (18-70 years old) with breast lesions were selected for this study. The Ivenia ABUS 2.0 (General Electric Company, USA) was used, with a probe frequency of 6-15 MHz. Adobe Photoshop CS6 software was used to calculate the pixel ratio of each ABUS image, and to draw an outline of the tumor cross-section. The resulting area (in pixels) was multiplied by the pixel ratio to yield the area of the tumor cross-section. The Wilcoxon signed rank test and Bland-Altman plot were used to compare mean differences and mean values, respectively, between the two methods.

Results:

There was no significant difference between the tumor volumes calculated by pixel method as compared to the traditional method (P>0.05). Repeated measurements of the same tumor volume were more consistent with the pixel method.

Conclusion:

The new pixel method is feasible for measuring breast tumor volume and has good validity and measurement stability.