Evaluation of a new method of calculating breast tumor volume based on automated breast ultrasound.
Front Oncol
; 12: 895575, 2022.
Article
em En
| MEDLINE
| ID: mdl-36176389
ABSTRACT
Objective:
To evaluate the effectiveness and advantages of a new method for calculating breast tumor volume based on an automated breast ultrasound system (ABUS).Methods:
A total of 42 patients (18-70 years old) with breast lesions were selected for this study. The Ivenia ABUS 2.0 (General Electric Company, USA) was used, with a probe frequency of 6-15 MHz. Adobe Photoshop CS6 software was used to calculate the pixel ratio of each ABUS image, and to draw an outline of the tumor cross-section. The resulting area (in pixels) was multiplied by the pixel ratio to yield the area of the tumor cross-section. The Wilcoxon signed rank test and Bland-Altman plot were used to compare mean differences and mean values, respectively, between the two methods.Results:
There was no significant difference between the tumor volumes calculated by pixel method as compared to the traditional method (P>0.05). Repeated measurements of the same tumor volume were more consistent with the pixel method.Conclusion:
The new pixel method is feasible for measuring breast tumor volume and has good validity and measurement stability.