Compatibility levels between blood gas analysis and central laboratory hemoglobin and electrolyte tests in pediatric patients: A single-center experience.
Konuksever, Dilek; Yucel, Sevinc Puren; Bölük, Oguz; Kiliç, Betül Orhan; Tasar, Medine Aysin.
Afiliação
  • Konuksever D; Department of Pediatrics, Turkish Ministry of Health Ankara City Hospital, Ankara, Turkey.
  • Yucel SP; Department of Biostatistics, Cukurova University, Adana, Turkey.
  • Bölük O; Department of Pediatrics, Ankara Training and Research Hospital, Ankara, Turkey.
  • Kiliç BO; Department of Pediatrics, Baskent University Hospital, Ankara, Turkey.
  • Tasar MA; Department of Pediatrics, Ankara Training and Research Hospital, Ankara, Turkey.
Paediatr Anaesth ; 33(2): 107-113, 2023 02.
Article em En | MEDLINE | ID: mdl-36178754
ABSTRACT

INTRODUCTION:

We aimed to evaluate the interchangeability of sodium, potassium, hemoglobin, and hematocrit measurement between the blood gas analyzers and laboratory automatic analyzers results.

METHODS:

This was a retrospective cross-sectional study. The results of 1927 paired samples analyzed simultaneously with the blood gas analyzer and the laboratory automatic analyzer were compared. The Bland-Altman and Cohen's kappa statistic detected the agreement between the two analyses.

RESULTS:

The limits of agreement (±1.96 standard deviation of the mean difference) were -11.1 to 20.3 for sodium, -1.9 to 0.5 for potassium, -16.1 to 12.9 for hematocrit, and -5.0 to 4.0 for hemoglobin. Agreement between the two analyses was not acceptable within the defined clinically acceptable limits. In addition, none of the kappa values were higher than 0.60, which highlights the lack of agreement between the two analyzers.

CONCLUSION:

The blood gas analyzers and laboratory automatic analyzers results cannot be used interchangeably.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas Automatizados de Assistência Junto ao Leito / Eletrólitos Tipo de estudo: Observational_studies Limite: Child / Humans Idioma: En Revista: Paediatr Anaesth Assunto da revista: ANESTESIOLOGIA / PEDIATRIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Turquia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Sistemas Automatizados de Assistência Junto ao Leito / Eletrólitos Tipo de estudo: Observational_studies Limite: Child / Humans Idioma: En Revista: Paediatr Anaesth Assunto da revista: ANESTESIOLOGIA / PEDIATRIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Turquia