Accuracy and Precision of a Novel Photogate System to Measure Toe Clearance on Stairs.
Skervin, Timmion K; Thomas, Neil M; Schofield, Andrew J; Hollands, Mark A; Maganaris, Constantinos N; O'Brien, Thomas D; Baltzopoulos, Vasilios; Foster, Richard J.
Afiliação
  • Skervin TK; Research to Improve Stair Climbing Safety (RISCS), Faculty of Science, School of Sport and Exercise Sciences, Liverpool John Moores University, Byrom Street, Liverpool L3 3AF, UK.
  • Thomas NM; Research to Improve Stair Climbing Safety (RISCS), Faculty of Science, School of Sport and Exercise Sciences, Liverpool John Moores University, Byrom Street, Liverpool L3 3AF, UK.
  • Schofield AJ; School of Psychology & Aston Research Centre for Healthy Ageing, College of Health and Life Sciences, Aston University, Birmingham B4 7ET, UK.
  • Hollands MA; Research to Improve Stair Climbing Safety (RISCS), Faculty of Science, School of Sport and Exercise Sciences, Liverpool John Moores University, Byrom Street, Liverpool L3 3AF, UK.
  • Maganaris CN; Research to Improve Stair Climbing Safety (RISCS), Faculty of Science, School of Sport and Exercise Sciences, Liverpool John Moores University, Byrom Street, Liverpool L3 3AF, UK.
  • O'Brien TD; Research to Improve Stair Climbing Safety (RISCS), Faculty of Science, School of Sport and Exercise Sciences, Liverpool John Moores University, Byrom Street, Liverpool L3 3AF, UK.
  • Baltzopoulos V; Research to Improve Stair Climbing Safety (RISCS), Faculty of Science, School of Sport and Exercise Sciences, Liverpool John Moores University, Byrom Street, Liverpool L3 3AF, UK.
  • Foster RJ; Research to Improve Stair Climbing Safety (RISCS), Faculty of Science, School of Sport and Exercise Sciences, Liverpool John Moores University, Byrom Street, Liverpool L3 3AF, UK.
Sensors (Basel) ; 23(5)2023 Feb 22.
Article em En | MEDLINE | ID: mdl-36904633
ABSTRACT

Background:

Toe clearance on stairs is typically measured using optoelectronic systems, though these are often constrained to the laboratory, due to their complex setups. Here we measured stair toe clearance through a novel prototype photogate setup and compared this to optoelectronic measurements.

Methods:

Twelve participants (age 22 ± 3 years) completed 25 stair ascent trials, each on a seven-step staircase. Toe clearance over the fifth step edge was measured using Vicon and the photogates. Twenty-two photogates were created in rows through laser diodes and phototransistors. The height of the lowest photogate broken at step-edge crossing was used to determine photogate toe clearance. A limits of agreement analysis and Pearson's correlation coefficient compared the accuracy, precision and relationship between systems.

Results:

We found a mean difference of -1.5 mm (accuracy) between the two measurement systems, with upper and lower limits (precision) of 10.7 mm and -13.8 mm, respectively. A strong positive correlation was also found (r = 70, n = 12, p = 0.009) between the systems.

Discussion:

The results suggest that photogates could be an option for measuring real-world stair toe clearances, where optoelectronic systems are not routinely used. Improvements to the design and measurement factors may help to improve the precision of the photogates.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Sensors (Basel) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Reino Unido