Impacted bone allograft personalised by a novel 3D printed customization kit produces high surgical accuracy in medial opening wedge high tibial osteotomy: a pilot study.
Van Genechten, Wouter; Van Haver, Annemieke; Bartholomeeusen, Stijn; Claes, Toon; Van Beek, Nathalie; Michielsen, Jozef; Claes, Steven; Verdonk, Peter.
Afiliação
  • Van Genechten W; Orthopedic Department, University Hospital Antwerp, Antwerp, Belgium. Wouter.vangenechten22@gmail.com.
  • Van Haver A; More Institute, Antwerp, Belgium. Wouter.vangenechten22@gmail.com.
  • Bartholomeeusen S; Orthopedic Department, AZ Herentals, Herentals, Belgium. Wouter.vangenechten22@gmail.com.
  • Claes T; More Institute, Antwerp, Belgium.
  • Van Beek N; Orthopedic Department, AZ Herentals, Herentals, Belgium.
  • Michielsen J; Orthopedic Department, AZ Herentals, Herentals, Belgium.
  • Claes S; Orthopedic Department, AZ Herentals, Herentals, Belgium.
  • Verdonk P; Orthopedic Department, University Hospital Antwerp, Antwerp, Belgium.
J Exp Orthop ; 10(1): 24, 2023 Mar 14.
Article em En | MEDLINE | ID: mdl-36917322
ABSTRACT

PURPOSE:

Contemporary medial opening wedge high tibial osteotomy (MOWHTO) still seems to struggle with inconsistent accuracy outcomes. Our objective was to assess surgical accuracy and short-term clinical outcomes when using 3D planning and a patient-specific instrumentation (PSI) kit to prepare customized bone allografts.

METHODS:

Thirty subjects (age 48y ± 13) were included in a double-center prospective case series. A low-dose CT-scan was performed to generate 3D bone models, a MOWHTO was simulated, and PSI was designed and 3D printed based on the complementary negative of the planned osteotomy gap. Clinical outcome was assessed at two, four, 12 weeks and one year using NRS, KOOS, UCLA activity score, EQ-5D and anchor questions. A linear-mixed model approach was implemented for data analysis.

RESULTS:

Preoperative 3D values were 175.0° ± 2.2 mechanical tibiofemoral angle (mTFA), 85.0° ± 3.0 medial proximal tibial angle (MPTA), and 94.1° ± 3.4 medial posterior tibial slope (MPTS). Target planning ranged from slight varus to the lateral tibial spine (slight valgus). Postoperative 3D analysis showed an accuracy of 1.1° ± 0.7 ΔMPTA (p = 0.04) and 1.2° ± 1.2 ΔMPTS (p = 0.11). NRS decreased from baseline 6.1 ± 1.9 to 2.7 ± 1.9 at four weeks (p < 0.001) and 1.7 ± 1.9 at one year (p < 0.001). KOOS increased from 31.4 ± 17.6 to 50.6 ± 20.6 at 12 weeks (p < 0.001) and to 71.8 ± 15.6 at one year (p < 0.001).

CONCLUSION:

The study suggests that 3D printed instrumentation to personalize structural bone allograft is a viable alternative method in MOWHTO that has the benefit of optimizing surgical accuracy while providing early and consistent pain relief after surgery.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Prognostic_studies Idioma: En Revista: J Exp Orthop Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Bélgica