Radioiodine-131 Therapy Used for Differentiated Thyroid Cancer Can Impair Titanium Dental Implants: An In Vitro Analysis.
Piciu, Doina; Bran, Simion; Moldovan, Marioara; Varvara, Simona; Piciu, Andra; Cuc, Stanca; Moisescu-Goia, Cristina; Barbus, Elena; Mester, Alexandru; Onisor, Florin.
Afiliação
  • Piciu D; Department of Endocrine Tumors and Nuclear Medicine, Institute of Oncology "Ion Chiricuta", 400015 Cluj-Napoca, Romania.
  • Bran S; Doctoral School, University of Medicine and Pharmacy "Iuliu Hatieganu", 400012 Cluj-Napoca, Romania.
  • Moldovan M; Department of Maxillofacial Surgery and Implantology, University of Medicine and Pharmacy "Iuliu Hatieganu", 400012 Cluj-Napoca, Romania.
  • Varvara S; Institute of Chemistry "Raluca Ripan", University Babes-Bolyai, 400294 Cluj-Napoca, Romania.
  • Piciu A; Faculty of Exact Sciences and Engineering, "1 Decembrie 1918" Alba Iulia University, 510009 Alba Iulia, Romania.
  • Cuc S; Department of Medical Oncology, University of Medicine and Pharmacy "Iuliu Hatieganu", 400012 Cluj-Napoca, Romania.
  • Moisescu-Goia C; Institute of Chemistry "Raluca Ripan", University Babes-Bolyai, 400294 Cluj-Napoca, Romania.
  • Barbus E; Department of Endocrine Tumors and Nuclear Medicine, Institute of Oncology "Ion Chiricuta", 400015 Cluj-Napoca, Romania.
  • Mester A; Department of Endocrine Tumors and Nuclear Medicine, Institute of Oncology "Ion Chiricuta", 400015 Cluj-Napoca, Romania.
  • Onisor F; Department of Oral Health, University of Medicine and Pharmacy "Iuliu Hatieganu", 400012 Cluj-Napoca, Romania.
Cancers (Basel) ; 15(9)2023 Apr 29.
Article em En | MEDLINE | ID: mdl-37174023
ABSTRACT

BACKGROUND:

The aim was to assess, in vitro, the effects of radioiodine-131 (I-131) on the structure of titanium implants. MATERIAL AND

METHODS:

A total of 28 titanium implants were divided into 7 groups (n = 4) and irradiated at 0, 6, 12, 24, 48, 192 and 384 hours. At the end of the experiment, each sample was investigated via scanning electron microscopy (SEM) and electrochemical measures.

RESULTS:

The control sample revealed a smooth and compact surface. The small micro-sized porosity is slightly visible at the macroscopic level, but the precise details cannot be observed. A mild exposure to the radioactive solution for 6 to 24 h showed a good preservation of the macro-structural aspects such as thread details and surface quality. Significant changes occurred after 48 h of exposure. It was noticed that the open-circuit potential (OCP) value of the non-irradiated implants move toward more noble potentials during the first 40 min of exposure to the artificial saliva and then stabilizes at a constant value of -143 mV. A displacement of the OCP values toward more negative values was observed for all irradiated implants; these potential shifts are decreasing, as the irradiation period of the tested implants increased.

CONCLUSION:

After exposure to I-131, the structure of titanium implants is well preserved up to 12 h. The eroded particles start to appear in the microstructural details after 24 h of exposure and their numbers progressively increase up to 384 h after exposure.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Cancers (Basel) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Romênia