High degree of alignment precision associated with total knee arthroplasty performed using a surgical robot or handheld navigation.
Shen, Tony S; Uppstrom, Tyler J; Walker, Paul J; Yu, Jonathan S; Cheng, Ryan; Mayman, David J; Jerabek, Seth A; Ast, Michael P.
Afiliação
  • Shen TS; Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA. shent@hss.edu.
  • Uppstrom TJ; Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA.
  • Walker PJ; Department of Orthopaedic Surgery, UCLA, Los Angeles, CA, USA.
  • Yu JS; Department of Orthopaedic Surgery, UCLA, Los Angeles, CA, USA.
  • Cheng R; Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA.
  • Mayman DJ; Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA.
  • Jerabek SA; Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA.
  • Ast MP; Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA.
Knee Surg Sports Traumatol Arthrosc ; 31(11): 4735-4740, 2023 Nov.
Article em En | MEDLINE | ID: mdl-37382709
ABSTRACT

PURPOSE:

The purpose of this study was to compare the precision of bony resections during total knee arthroplasty (TKA) performed using different computer-assisted technologies.

METHODS:

Patients who underwent a primary TKA using an imageless accelerometer-based handheld navigation system (KneeAlign2®, OrthAlign Inc.) or computed tomography-based large-console surgical robot (Mako®, Stryker Corp.) from 2017 to 2020 were retrospectively reviewed. Templated alignment targets and demographic data were collected. Coronal plane alignment of the femoral and tibial components and tibial slope were measured on postoperative radiographs. Patients with excessive flexion or rotation preventing accurate measurement were excluded.

RESULTS:

A total of 240 patients who underwent TKA using either a handheld (n = 120) or robotic (n = 120) system were included. There were no statistically significant differences in age, sex, and BMI between groups. A small but statistically significant difference in the precision of the distal femoral resection was observed between the handheld and robotic cohorts (1.5° vs. 1.1° difference between templated and measured alignments, p = 0.024), though this is likely clinically insignificant. There were no significant differences in the precision of the tibial resection between the handheld and robotic groups (coronal plane 0.9° vs. 1.0°, n.s.; sagittal plane 1.2° vs. 1.1°, n.s.). There were no significant differences in the rate of overall precision between cohorts (n.s.).

CONCLUSIONS:

A high degree of component alignment precision was observed for both imageless handheld navigation and CT-based robotic cohorts. Surgeons considering options for computer-assisted TKA should take other important factors, including surgical principles, templating software, ligament balancing, intraoperative adjustability, equipment logistics, and cost, into account. LEVEL OF EVIDENCE III.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Risk_factors_studies Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Assunto da revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos