Dual Mobility Articulation in Total Hip Arthroplasty: Mixed Femoral and Acetabular Components are a Feasible Option.
Holland, Christopher; Cochrane, Niall; Hinton, Zoe; Wellman, Samuel; Seyler, Thorsten; Bolognesi, Michael; Ryan, Sean.
Afiliação
  • Holland C; Campbell Clinic Orthopaedics, University of Tennessee Health Science Center, College of Medicine, Department of Orthopaedic Surgery and Biomedical Engineering.
  • Cochrane N; Department of Orthopaedic Surgery, Duke University Medical Center, Durham, North Carolina.
  • Hinton Z; Department of Orthopaedic Surgery, Duke University Medical Center, Durham, North Carolina.
  • Wellman S; Department of Orthopaedic Surgery, Duke University Medical Center, Durham, North Carolina.
  • Seyler T; Department of Orthopaedic Surgery, Duke University Medical Center, Durham, North Carolina.
  • Bolognesi M; Department of Orthopaedic Surgery, Duke University Medical Center, Durham, North Carolina.
  • Ryan S; Department of Orthopaedic Surgery, Duke University Medical Center, Durham, North Carolina.
J Arthroplasty ; 2024 Jan 14.
Article em En | MEDLINE | ID: mdl-38220027
ABSTRACT

BACKGROUND:

The utilization of a different manufacturer for the prosthetic femoral head and the polyethylene insert in dual mobility (DM) for total hip arthroplasty (THA) may be necessary, especially in the revision setting. However, there is no data in the literature about this application. This study evaluated the outcomes of mixed manufacturer components, with the hypothesis that there would be no difference in measured outcomes compared to matched components.

METHODS:

The DM articulations implanted during THA revision were retrospectively reviewed from 2011 to 2017. The study group was then stratified into 2 cohorts matching components or mixed components. Of 130 hips included in the study with DM articulations with average follow-up of 7 years, 103 had mixed and 27 had matching manufacturer components. Rates of all cause reoperation and revision, intraprosthetic dislocation, dislocation, and aseptic loosening were compared using Chi-squared and Fisher's exact test; survival analysis was also performed.

RESULTS:

Matched and mixed manufacturer implants had no significant difference between all cause reoperation (33 versus 25.2%), dislocation (14.8 versus 7.7%), and aseptic loosening (3.7 versus 3.9%), respectively. Higher rates of intraprosthetic dislocation (11 versus 0.97%) were observed in the matching component cohort. Survival analysis showed similar outcomes at 2, 5, and 10 years.

CONCLUSIONS:

Mixed-component DM articulations show similar results compared to matching components. The off-label use of mixed manufacture DM articulation in THA is a feasible and safe option in the correct patient. Furthermore, when encountering a well-fixed femoral stem or acetabular shell, the use of a mixed component DM articulations may reduce the morbidity for the patient and prevent revision of all components.