Evaluation of a novel, low-cost, 3D printed video laryngoscope with borescope in anesthetized Beagle dogs.
Son, Won-Gyun; Sung, Taehoon; Shin, Donghwi; Rhee, Suehyung; Nam, Changhoon; Kim, Minha; Park, Chailin; Lee, Jungha; Kim, Junsoo; Lee, Inhyung.
Afiliação
  • Son WG; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea. Electronic address: carpeego@snu.ac.kr.
  • Sung T; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea.
  • Shin D; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea.
  • Rhee S; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea.
  • Nam C; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea.
  • Kim M; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea.
  • Park C; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea.
  • Lee J; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea.
  • Kim J; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea.
  • Lee I; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Seoul National University, Seoul, Republic of Korea.
Vet Anaesth Analg ; 51(3): 266-270, 2024.
Article em En | MEDLINE | ID: mdl-38565449
ABSTRACT

OBJECTIVE:

To develop and evaluate a low-cost three-dimensional (3D)-printed video laryngoscope (VLVET) for use with a commercial borescope. STUDY

DESIGN:

Instrument development and pilot study. ANIMALS A total of six adult male Beagle dogs.

METHODS:

The VLVET consisted of a laryngoscope handle and a Miller-type blade, and a detachable camera holder that attached to various locations along the blade. The laryngoscope and camera holder were 3D-printed using black polylactic acid filament. Dogs were premedicated with intravenous (IV) medetomidine (15 µg kg-1) and anesthesia induced with IV alfaxalone (1.5 mg kg-1). The VLVET, combined with a borescope, was used for laryngeal visualization and intubation. Performance was evaluated by comparing direct and video-assisted views in sternal recumbency. The borescope camera was sequentially positioned at 2, 4, 6, 8 and 10 cm from the blade tip (distanceLARYNX-CAM), which was placed on the epiglottis during intubation or laryngoscopy. At the 10 cm distanceLARYNX-CAM, laryngeal visualization was sequentially scored at inter-incisor gaps of 10, 8, 6, 4 and 2 cm. Laryngeal visualization scores (0-3 range, with 0 = obstructed and 3 = unobstructed views) were statistically analyzed using the Friedman's test.

RESULTS:

Under direct visualization, the 2 cm distanceLARYNX-CAM had a significantly lower score compared with all other distanceLARYNX-CAM (all p = 0.014) because the view was obstructed by the camera holder and borescope camera. With both direct and camera-assisted views, visualization scores were higher at inter-incisor gaps ≥ 4 cm compared with 2 cm (all p < 0.05). CONCLUSIONS AND CLINICAL RELEVANCE During laryngoscopy and intubation, the VLVET and borescope facilitated both direct and video laryngoscopy at distanceLARYNX-CAM in Beagle dogs when inter-incisor gaps were ≥ 4 cm.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Laringoscópios / Impressão Tridimensional / Intubação Intratraqueal Limite: Animals Idioma: En Revista: Vet Anaesth Analg Assunto da revista: ANESTESIOLOGIA / MEDICINA VETERINARIA Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Laringoscópios / Impressão Tridimensional / Intubação Intratraqueal Limite: Animals Idioma: En Revista: Vet Anaesth Analg Assunto da revista: ANESTESIOLOGIA / MEDICINA VETERINARIA Ano de publicação: 2024 Tipo de documento: Article