His potential injury as the end point of screwing by a continuous recording technique in His bundle pacing: A case report.
Zhong, Jinyan; Jiang, Longfu.
Afiliação
  • Zhong J; Department of Cardiology, Ningbo No. 2 hospital, Ningbo, Zhejiang, China.
  • Jiang L; Department of Cardiology, Ningbo No. 2 hospital, Ningbo, Zhejiang, China.
Pacing Clin Electrophysiol ; 47(6): 771-775, 2024 Jun.
Article em En | MEDLINE | ID: mdl-38576198
ABSTRACT

BACKGROUND:

His bundle pacing (HBP) engaged electrical activation of both ventricles by stimulating the His-Purkinje network, which could avoid marked ventricles dyssynchrony. The lead was given three to five clockwise rotations at the site with the His potential to anchor the interventricular septum. In 2018, the Multicenter His Bundle Pacing Collaborative Working Group recommended that the His bundle capture threshold should be lower than 2.5 V/1 ms in non-pacing-dependent patients, and pacing-dependent patients should have a lower adjacent ventricular capture threshold as self-backup. Therefore, to avoid safety issues such as loss of capture caused by increased threshold, we believe that more stringent criteria should be adopted in patients with atrioventricular block (AVB). In previous studies, the connection cable needed to be disconnected during the screwing. When the procedure was finished, the performer found that the patients with His bundle injury could obtain a lower threshold than those without His bundle injury. Although no studies of new bundle branch block (BBB) or AVB by the acute His bundle injury was reported. However, It is worrying that the damage of His bundle seems random during the procedure. How to balance avoiding severe injury with a lower capture threshold? At present, we report a case of light His injury and lower His capture threshold under continuous intracardiac electrocardiogram monitoring.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fascículo Atrioventricular / Estimulação Cardíaca Artificial / Eletrocardiografia Limite: Aged / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fascículo Atrioventricular / Estimulação Cardíaca Artificial / Eletrocardiografia Limite: Aged / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China