Septal venous channel perforation during left bundle branch area pacing: a prospective study.
Europace
; 26(6)2024 Jun 03.
Article
em En
| MEDLINE
| ID: mdl-38703372
ABSTRACT
AIMS:
To characterize the diagnosis, frequency, and procedural implications of septal venous channel perforation during left bundle branch area pacing (LBBAP). METHODS ANDRESULTS:
All consecutive patients undergoing LBBAP over an 8-month period were prospectively studied. During lead placement, obligatory septal contrast injection was performed twice, at initiation (implant entry zone) and at completion (fixation zone). An intuitive fluoroscopic schema using orthogonal views (left anterior oblique/right anterior oblique) and familiar landmarks is described. Using this, we resolved zonal distribution (I-VI) of lead position on the ventricular septum and its angulation (post-fixation angle θ). Subjects with and without septal venous channel perforation were compared. Sixty-one patients {male 57.3%, median age [interquartile range (IQR)] 69.5 [62.5-74.5] years} were enrolled. Septal venous channel perforation was observed in eight (13.1%) patients [male 28.5%, median age (IQR) 64 (50-75) years]. They had higher frequency of (i) right-sided implant (25% vs. 1.9%, P = 0.04), (ii) fixation in zone III at the mid-superior septum (75% vs. 28.3%, P = 0.04), (iii) steeper angle of fixation-median θ (IQR) [19 (10-30)° vs. 5 (4-19)°, P = 0.01], and (iv) longer median penetrated-lead length (IQR) [13 (10-14.8) vs. 10 (8.5-12.5) mm, P = 0.03]. Coronary sinus drainage of contrast was noted in five (62.5%) patients. Abnormal impedance drops during implantation (12.5% vs. 5.7%, P = NS) were not significantly different.CONCLUSION:
When evaluated systematically, septal venous channel perforation may be encountered commonly after LBBAP. The fiducial reference framework described using fluoroscopic imaging identified salient associated findings. This may be addressed with lead repositioning to a more inferior location and is not associated with adverse consequence acutely or in early follow-up.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Estimulação Cardíaca Artificial
Limite:
Aged
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Europace
Assunto da revista:
CARDIOLOGIA
/
FISIOLOGIA
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Índia