Influence of clinical expertise and practical experience on transfer accuracy in guided dental implant placement - an in vitro study.
Reiff, Florian Sebastian; Bischoff, Charlotte; Woelfler, Henriette; Roehling, Stefan.
Afiliação
  • Reiff FS; Center of Oral Health, University of Greifswald, Greifswald, Germany. Florian.Sebastian.Reiff@stud.dtmd.eu.
  • Bischoff C; Straumann Group, Freiburg im Breisgau, Germany. Florian.Sebastian.Reiff@stud.dtmd.eu.
  • Woelfler H; Straumann Group, Basel, Switzerland.
  • Roehling S; University of Bamberg, Bamberg, Germany.
Oral Maxillofac Surg ; 28(4): 1491-1500, 2024 Dec.
Article em En | MEDLINE | ID: mdl-38914846
ABSTRACT

PURPOSE:

To investigate whether inexperienced users applying a static navigation system can perform in-vitro a fully guided implant placement protocol and achieve similar results in terms of accuracy compared to experienced clinicians.

METHODS:

Based on 36 identical resin models, a computer-assisted implant planning was performed and a surgical guide was produced accordingly. Three study groups were composed with 12 operators, each control group with experienced surgeons (DOC), test group 1 with dental technicians (TEC) and test group 2 with non-specialists (OFC). Using a fully guided drilling protocol, two implants were placed into each of the 36 models. Subsequently, the differences between the virtually planned and final implant positions were determined and the transfer accuracy was evaluated.

RESULTS:

For the control group DOC, the mean value of axial deviation was 1.90 ± 1.15 degrees, for 3-dimensional deviation at the implant base 0.52 ± 0.33 mm, for 3-dimensional deviation at the implant tip 0.76 ± 0.39 mm and for vertical deviation at the implant tip - 0.11 ± 0.51 mm. For corresponding parameters, the mean values of test group TEC were 1.99 ± 0.87 degrees, 0.42 ± 0.21 mm, 0.68 ± 0.30 mm and - 0.03 ± 0.33 mm and for test group OFC 2.29 ± 1.17 degrees, 0.63 ± 0.35 mm, 0.89 ± 0.43 mm and - 0.24 ± 0.57 mm, respectively. The results did not reveal any statistically significant differences between the control and the 2 test groups (p˃0.05).

CONCLUSION:

The results of the present in-vitro study demonstrated that inexperienced users applying a static navigation system can perform a fully guided implant placement protocol and achieve similar results in terms of accuracy compared to experienced clinicians in this specific in vitro setup.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Competência Clínica / Cirurgia Assistida por Computador / Implantação Dentária Endóssea Limite: Humans Idioma: En Revista: Oral Maxillofac Surg Assunto da revista: ODONTOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Competência Clínica / Cirurgia Assistida por Computador / Implantação Dentária Endóssea Limite: Humans Idioma: En Revista: Oral Maxillofac Surg Assunto da revista: ODONTOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Alemanha