Acceptance testing computerized radiation therapy treatment planning systems: direct utilization of CT scan data.
Med Phys ; 12(2): 237-42, 1985.
Article em En | MEDLINE | ID: mdl-4000085
ABSTRACT
The availability of computerized radiation therapy treatment planning systems that utilize computed tomography (CT) scan data requires testing additional to that routinely needed for non-CT systems. These additional items include dimensioning verification, establishing CT number-to-tissue property conversions, verifying the accuracy of heterogeneity corrected dose predictions and autocontouring. One testing protocol is presented and sample results from an Atomic Energy of Canada Theraplan L system are presented.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radioterapia / Tomografia Computadorizada por Raios X Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 1985 Tipo de documento: Article