1.
Med Phys ; 39(12): 7272-90, 2012 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-23231278

RESUMO

PURPOSE: Incident learning plays a key role in improving quality and safety in a wide range of industries and medical disciplines. However, implementing an effective incident learning system is complex, especially in radiation oncology. One current barrier is the lack of technical standards to guide users or developers. This report, the product of an initiative by the Work Group on Prevention of Errors in Radiation Oncology of the American Association of Physicists in Medicine, provides technical recommendations for the content and structure of incident learning databases in radiation oncology. METHODS: A panel of experts was assembled and tasked with developing consensus recommendations in five key areas: definitions, process maps, severity scales, causality taxonomy, and data elements. Experts included representatives from all major North American radiation oncology organizations as well as users and developers of public and in-house reporting systems with over two decades of collective experience. Recommendations were developed that take into account existing incident learning systems as well as the requirements of outside agencies. RESULTS: Consensus recommendations are provided for the five major topic areas. In the process mapping task, 91 common steps were identified for external beam radiation therapy and 88 in brachytherapy. A novel feature of the process maps is the identification of "safety barriers," also known as critical control points, which are any process steps whose primary function is to prevent errors or mistakes from occurring or propagating through the radiotherapy workflow. Other recommendations include a ten-level medical severity scale designed to reflect the observed or estimated harm to a patient, a radiation oncology-specific root causes table to facilitate and regularize root-cause analyses, and recommendations for data elements and structures to aid in development of electronic databases. Also presented is a list of key functional requirements of any reporting system. CONCLUSIONS: Incident learning is recognized as an invaluable tool for improving the quality and safety of treatments. The consensus recommendations in this report are intended to facilitate the implementation of such systems within individual clinics as well as on broader national and international scales.


Assuntos
Sistemas de Gerenciamento de Base de Dados/normas , Bases de Dados Factuais/normas , Notificação de Abuso , Lesões por Radiação/epidemiologia , Radioterapia (Especialidade)/normas , Sistema de Registros/normas , Gestão de Riscos/normas
2.
Dis Esophagus ; 21(5): 437-44, 2008.
Artigo em Inglês | MEDLINE | ID: mdl-19125798

RESUMO

Locoregional relapse occurs in over half of gastric cancer patients who undergo potentially curative resection. Adjuvant chemoradiation reduces locoregional relapse, but often requires irradiating large fields and is limited by poor patient tolerance. This study explores the potential dosimetric benefit in reducing the radiation dose to normal structures by treating gastroesophageal (GE) junction/proximal gastric cancers with preoperative rather than adjuvant radiotherapy. Five cases of GE junction/proximal gastric cancer patients treated postoperatively with curative intent were selected. The actual target contours were then modified to reflect hypothetical target volumes which would have been used had the patients been treated preoperatively. Hypothetical preoperative treatment plans were generated for each patient based on these modified contours. The hypothetical preoperative treatment plans were then compared to the actual postoperative plans with respect to dose-volume parameters including lung mean dose, lung V20, heart V20 and V30, and mean doses to abdominal structures. Target volumes were smaller with preoperative treatment, with an average reduction of 23%. Comparative dose-volume histogram (DVH) analysis showed the resultant composite lung doses were reduced in the preoperative plans by 50-79%. In all patients, the proportion of lungs receiving at least 20 Gy (V20) was substantially reduced using preoperative treatment (1.9% vs. 9.7% in the 3-D conformal patient; mean of 3.1% vs. 17.6% in the intensity modulated radiation therapy patients). Likewise, the volume of heart receiving at least 30 Gy was dramatically reduced in all preoperative plans (15.8% vs. 35.4%). Doses to the kidneys, liver and spinal cord were comparable in both approaches. Preoperative treatment of GE junction and proximal gastric cancer patients offers the potential to decrease the radiation dose received by normal thoracic structures.


Assuntos
Adenocarcinoma/radioterapia , Junção Esofagogástrica/patologia , Junção Esofagogástrica/efeitos da radiação , Recidiva Local de Neoplasia/prevenção & controle , Neoplasias Gástricas/diagnóstico por imagem , Adenocarcinoma/mortalidade , Adenocarcinoma/patologia , Adenocarcinoma/cirurgia , Idoso , Junção Esofagogástrica/cirurgia , Feminino , Seguimentos , Gastrectomia/métodos , Humanos , Masculino , Pessoa de Meia-Idade , Invasividade Neoplásica/patologia , Recidiva Local de Neoplasia/mortalidade , Estadiamento de Neoplasias , Cuidados Pós-Operatórios , Cuidados Pré-Operatórios/métodos , Probabilidade , Radiografia , Dosagem Radioterapêutica , Radioterapia Conformacional/métodos , Radioterapia de Intensidade Modulada/métodos , Valores de Referência , Medição de Risco , Estudos de Amostragem , Neoplasias Gástricas/mortalidade , Neoplasias Gástricas/patologia , Neoplasias Gástricas/cirurgia , Análise de Sobrevida , Fatores de Tempo , Resultado do Tratamento
3.
Clin Oncol (R Coll Radiol) ; 29(9): 562-567, 2017 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-28739028

RESUMO

AIMS: To discuss the background for incident reporting and learning systems, as well as the infrastructure and operational aspects to run them. MATERIALS AND METHODS: Information from peer-reviewed literature, online resources and the authors' experience synthesised into a concise understanding of the topic. RESULTS: Incident learning systems can be local, national or international, each having the same basic goals but facilitating different audiences and environments. A key component of any reporting and learning system is timely and effective analysis of near-misses and incidents as well as feedback to the users of the system. It is important for staff to know that reports are acknowledged, analysed and acted upon. There is a need to comply with current European legislation and other national systems, which can be addressed together with the steps required for comprehensive management of an incident. CONCLUSION: Reporting and learning from incidents and near-misses is a key component of quality and safety in radiotherapy. A major benefit of the national or international systems is the potential for a larger database of incidents, supporting wider analysis and comparison, and sharing of knowledge across a larger community.


Assuntos
Radioterapia (Especialidade)/métodos , Gestão de Riscos/métodos , Humanos
4.
Phys Med Biol ; 50(5): 891-907, 2005 Mar 07.
Artigo em Inglês | MEDLINE | ID: mdl-15798263

RESUMO

This paper reviews the effect of statistical uncertainties on radiotherapy treatment planning using Monte Carlo simulations. We discuss issues related to the statistical analysis of Monte Carlo dose calculations for realistic clinical beams using various variance reduction or time saving techniques. We discuss the effect of statistical uncertainties on dose prescription and monitor unit calculation for conventional treatment and intensity-modulated radiotherapy (IMRT) based on Monte Carlo simulations. We show the effect of statistical uncertainties on beamlet dose calculation and plan optimization for IMRT and other advanced treatment techniques such as modulated electron radiotherapy (MERT). We provide practical guidelines for the clinical implementation of Monte Carlo treatment planning and show realistic examples of Monte Carlo based IMRT and MERT plans.


Assuntos
Neoplasias de Cabeça e Pescoço/radioterapia , Processamento de Imagem Assistida por Computador/métodos , Aceleradores de Partículas/instrumentação , Planejamento da Radioterapia Assistida por Computador/métodos , Simulação por Computador , Relação Dose-Resposta à Radiação , Humanos , Modelos Estatísticos , Método de Monte Carlo , Distribuição Normal , Dosagem Radioterapêutica , Radioterapia Conformacional/métodos , Espalhamento de Radiação
5.
Int J Radiat Oncol Biol Phys ; 32(1): 159-64, 1995 Apr 30.
Artigo em Inglês | MEDLINE | ID: mdl-7721612

RESUMO

PURPOSE: To determine the effects of extended source-to-surface distance (SSD) on dose distributions for a range of medium energy electron beams and cone sizes. METHODS AND MATERIALS: The depth-dose curves and isodose distributions of 6 MeV, 10 MeV, and 14 MeV electron beams from a dual photon and multielectron energies linear accelerator were studied. To examine the influence of cone size, the smallest and the largest cone sizes available were used. Measurements were carried out in a water phantom with the water surface set at three different SSDs from 101 to 116 cm. RESULTS: In the region between the phantom surface and the depth of maximum dose, the depth-dose decreases as the SSD increases for all electron beam energies. The effects of extended SSD in the region beyond the depth of maximum dose are unobservable and, hence, considered minimal. Extended SSD effects are apparent for higher electron beam energy with small cone size causing the depth of maximum dose and the rapid dose fall-off region to shift deeper into the phantom. However, the change in the depth-dose curve is small. On the other hand, the rapid dose fall-off region is essentially unaltered when the large cone is used. The penumbra enlarges and electron beam flatness deteriorates with increasing SSD.


Assuntos
Elétrons , Modelos Anatômicos , Aceleradores de Partículas , Dosagem Radioterapêutica , Desenho de Equipamento
6.
Int J Radiat Oncol Biol Phys ; 34(4): 917-22, 1996 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-8598371

RESUMO

PURPOSE: To study the dose distributions of high dose rate (HDR) endobronchial implants using the dose nonuniformity ratio (DNR) and three volumetric irradiation indices. METHODS AND MATERIALS: Multiple implants were configured by allowing a single HDR 192Ir source to step through a length of 6 cm along an endobronchial catheter. Dwell times were computed to deliver a dose of 5 Gy to points 1 cm away from the catheter axis. Five sets of source configurations, each with different dwell position spacings from 0.5 to 3.0 cm, were evaluated. Three-dimensional (3D) dose distributions were then generated for each source configuration. Differential and cumulative dose-volume curves were generated to quantify the degree of target volume coverage, dose nonuniformity within the target volume, and irradiation of tissues outside the target volume. Evaluation of the implants were made using the DNR and three volumetric irradiation indices. RESULTS: The observed isodose distributions were not able to satisfy all the dose constraints. The ability to optimally satisfy the dose constraints depended on the choice of dwell position spacing and the specification of the dose constraint points. The DNR and irradiation indices suggest that small dwell position spacing does not result in a more homogeneous dose distribution for the implant. This study supports the existence of a relationship between the dwell position spacing and the distance from the catheter axis to the reference dose or dose constraint points. Better dose homogeneity for an implant can be obtained if the spacing of the dwell positions are about twice the distance from the catheter axis to the reference dose or dose constraint points.


Assuntos
Braquiterapia/métodos , Brônquios , Radioisótopos de Irídio/uso terapêutico , Dosagem Radioterapêutica
7.
Int J Radiat Oncol Biol Phys ; 28(4): 979-83, 1994 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-8138451

RESUMO

PURPOSE: Evaluating the validity of the universal wedge equation over the range of 60Cobalt (60CO) to 25 MV photon beam energies. METHODS AND MATERIALS: The universal wedge equation relates the beam weight of the wedged field to the effective wedge angle produced by combining the wedged field to the nonwedged field and is expressed as tan (theta) = B tan (theta w), where B is the normalized weight imposed on the wedged field, theta w is the maximum wedge angle of the wedge filter, and theta is the effective wedge angle. The isodose distributions from 60Co to 25 MV photon beam energies were used. For each photon energy, the isodose distributions of wedged and nonwedged fields were combined in different proportions. The effective wedge angle was determined from each resultant isodose distribution. RESULTS: The relationship between the weight of wedged field and the effective wedge angle was found to be nonlinear for wedge filters with maximum wedge angle greater than 30 degrees. The universal wedge equation predicts the effective wedge angle to within 2 degrees compared to the measured value.


Assuntos
Radioisótopos de Cobalto/uso terapêutico , Fótons , Dosagem Radioterapêutica , Humanos
8.
Int J Radiat Oncol Biol Phys ; 39(1): 31-7, 1997 Aug 01.
Artigo em Inglês | MEDLINE | ID: mdl-9300737

RESUMO

PURPOSE: A retrospective analysis to determine differences in survival of patients with pancreatic adenocarcinoma treated by radical surgery with and without adjuvant radiation therapy. METHODS AND MATERIALS: Between 1980 and 1995, 249 patients with pancreatic tumors were identified at the Medical College of Ohio. Forty-four of these patients underwent radical surgical procedures with curative intent. These patients were divided into four groups according to treatment: surgery alone (n = 14), surgery plus intraoperative radiation therapy (IORT) (n = 6), surgery plus external beam radiation therapy (EBRT) (n = 14), or surgery plus both IORT and EBRT (n = 10). Outcome and survival were analyzed among the four groups. RESULTS: The median survival time of patients treated with radical surgery alone was 6.5 months. The median survival time for the surgery plus IORT group was 9 months; however, 33.3% (two of six) of these patients survived longer than 5 years. This survival pattern was borderline significantly better than that for the surgery alone group (p = 0.0765). The surgery plus EBRT and the surgery plus IORT and EBRT groups had median survival times of 14.5 and 17.5 months, respectively. These were significantly better than that of the surgery alone group (p = 0.0004 and p = 0.0002, respectively). The addition of radiation therapy did not affect the treatment complication rate. CONCLUSION: The survival of patients who were treated with radical surgery alone was significantly poorer than that of patients who received adjuvant radiation therapy. These results are consistent with other studies in the literature. Patients treated with all three modalities (surgery, IORT, and EBRT) displayed the best median survival time.


Assuntos
Neoplasias Pancreáticas/radioterapia , Neoplasias Pancreáticas/cirurgia , Idoso , Idoso de 80 Anos ou mais , Causas de Morte , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Neoplasias Pancreáticas/tratamento farmacológico , Neoplasias Pancreáticas/mortalidade , Radioterapia Adjuvante , Estudos Retrospectivos , Análise de Sobrevida
9.
Med Phys ; 26(10): 2133-43, 1999 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-10535630

RESUMO

The purpose of this study was to implement the Monte Carlo method for clinical radiotherapy dose calculations. We used the EGS4/BEAM code to obtain the phase-space data for 6-20 MeV electron beams and 4, 6, and 15 MV photon beams for Varian Clinac 1800, 2100C, and 2300CD accelerators. A multiple-source model was used to reconstruct the phase-space data for both electron and photon beams, which retained the accuracy of the Monte Carlo beam data. The multiple-source model reduced the phase-space data storage requirement by a factor of 1000 and the accelerator simulation time by a factor of 10 or more. Agreement within 2% was achieved between the Monte Carlo calculations and measurements of the dose distributions in homogeneous and heterogeneous phantoms for various field sizes, source-surface distances, and beam modulations. The Monte Carlo calculated electron output factors were within 2% of the measured values for various treatment fields while the heterogeneity correction factors for various lung and bone phantoms were within 1% for photon beams and within 2% for electron beams. The EGS4/DOSXYZ Monte Carlo code was used for phantom and patient dose calculations. The results were compared to the dose distributions produced by a conventional treatment planning system and an intensity-modulated radiotherapy inverse-planning system. Significant differences (>5% in dose and >5 mm shift in isodose lines) were found between Monte Carlo calculations and the analytical calculations implemented in the commercial systems. Treatment sites showing the largest dose differences were for head and neck, lung, and breast cases.


Assuntos
Método de Monte Carlo , Planejamento da Radioterapia Assistida por Computador , Radioterapia/instrumentação , Radioterapia/métodos , Algoritmos , Apresentação de Dados , Humanos , Neoplasias Pulmonares/diagnóstico por imagem , Neoplasias Pulmonares/radioterapia , Neoplasias Nasofaríngeas/diagnóstico por imagem , Neoplasias Nasofaríngeas/radioterapia , Imagens de Fantasmas , Fótons , Dosagem Radioterapêutica , Software , Tomografia Computadorizada por Raios X/métodos
10.
Phys Med Biol ; 45(8): 2151-61, 2000 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-10958186

RESUMO

Dose-volume histograms (DVHs) of the dose distributions calculated by the Monte Carlo method contain statistical uncertainties. The Monte Carlo DVH can be considered as blurred from the noiseless DVH by the statistical uncertainty. The focus of the present work is on the removal of the statistical uncertainty effect on the Monte Carlo DVHs and the reconstruction of the noiseless DVHs. We first study the effect of statistical uncertainty. It is found that the steeper the DVH, the more significant the effect. For typical critical structure DVHs the effect is usually negligible. For the target DVHs the effect could be clinically significant, depending on the value of uncertainty and the slope of the DVH. We then propose an iterative reconstruction algorithm. Using the DVHs and statistical uncertainties from the Monte Carlo simulations, we are able to reconstruct the noiseless DVHs. A hypothetical example and a number of clinical cases have been used to test the proposed algorithm. For each clinical case, two Monte Carlo simulations (denoted A and B) were performed. Simulation A has very large statistical uncertainties (about 10% of dose in the target volume) while simulation B has very small uncertainties (about 1%). DVHs from simulation B were used to approximate the noiseless DVHs. Using the proposed algorithm, the effect of statistical uncertainty can be removed from the DVHs of simulation A. The reconstructed DVHs were in good agreement with the DVHs from simulation B. The proposed approach is expected to be useful in removing the blurring effect on a quickly calculated Monte Carlo DVH when performing the iterative forward treatment planning.


Assuntos
Método de Monte Carlo , Radiometria/métodos , Algoritmos , Simulação por Computador , Humanos , Modelos Estatísticos , Neoplasias/radioterapia , Planejamento da Radioterapia Assistida por Computador
11.
Phys Med Biol ; 45(10): 2969-85, 2000 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-11049183

RESUMO

A new EGS4/PRESTA Monte Carlo user code, MCDOSE, has been developed as a routine dose calculation tool for radiotherapy treatment planning. It is suitable for both conventional and intensity modulated radiation therapy. Two important features of MCDOSE are the inclusion of beam modifiers in the patient simulation and the implementation of several variance reduction techniques. Before this tool can be used reliably for clinical dose calculation, it must be properly validated. The validation for beam modifiers has been performed by comparing the dose distributions calculated by MCDOSE and the well-benchmarked EGS4 user codes BEAM and DOSXYZ. Various beam modifiers were simulated. Good agreement in the dose distributions was observed. The differences in electron cutout factors between the results of MCDOSE and measurements were within 2%. The accuracy of MCDOSE with various variance reduction techniques was tested by comparing the dose distributions in different inhomogeneous phantoms with those calculated by DOSXYZ without variance reduction. The agreement was within 1.0%. Our results demonstrate that MCDOSE is accurate and efficient for routine dose calculation in radiotherapy treatment planning, with or without beam modifiers.


Assuntos
Elétrons , Método de Monte Carlo , Planejamento da Radioterapia Assistida por Computador/métodos , Osso e Ossos/efeitos da radiação , Humanos , Pulmão/efeitos da radiação , Imagens de Fantasmas , Radioterapia Conformacional/métodos
12.
Phys Med Biol ; 48(7): 909-24, 2003 Apr 07.
Artigo em Inglês | MEDLINE | ID: mdl-12701895

RESUMO

Recently, energy- and intensity-modulated electron radiotherapy (MERT) has garnered a growing interest for the treatment of superficial targets. In this work. we carried out a comparative dosimetry study to evaluate MERT, photon beam intensity-modulated radiation therapy (IMRT) and conventional tangential photon beams for the treatment of breast cancer. A Monte Carlo based treatment planning system has been investigated, which consists of a set of software tools to perform accurate dose calculation, treatment optimization, leaf sequencing and plan analysis. We have compared breast treatment plans generated using this home-grown treatment optimization and dose calculation software forthese treatment techniques. The MERT plans were planned with up to two gantry angles and four nominal energies (6, 9, 12 and 16 MeV). The tangential photon treatment plans were planned with 6 MV wedged photon beams. The IMRT plans were planned using both multiple-gantry 6 MV photon beams or two 6 MV tangential beams. Our results show that tangential IMRT can reduce the dose to the lung, heart and contralateral breast compared to conventional tangential wedged beams (up to 50% reduction in high dose volume or 5 Gy in the maximum dose). MERT can reduce the maximum dose to the lung by up to 20 Gy and to the heart by up to 35 Gy compared to conventional tangential wedged beams. Multiple beam angle IMRT can significantly reduce the maximum dose to the lung and heart (up to 20 Gy) but it induces low and medium doses to a large volume of normal tissues including lung, heart and contralateral breast. It is concluded that MERT has superior capabilities to achieve dose conformity both laterally and in the depth direction, which will be well suited for treating superficial targets such as breast cancer.


Assuntos
Neoplasias da Mama/radioterapia , Fótons/uso terapêutico , Radiometria/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia Conformacional/métodos , Radioterapia de Alta Energia/métodos , Simulação por Computador , Dosagem Radioterapêutica
13.
Phys Med Biol ; 45(2): 411-27, 2000 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-10701512

RESUMO

Photon beams of 4, 6 and 15 MV from Varian Clinac 2100C and 2300C/D accelerators were simulated using the EGS4/BEAM code system. The accelerators were modelled as a combination of component modules (CMs) consisting of a target, primary collimator, exit window, flattening filter, monitor chamber, secondary collimator, ring collimator, photon jaws and protection window. A full phase space file was scored directly above the upper photon jaws and analysed using beam data processing software, BEAMDP, to derive the beam characteristics, such as planar fluence, angular distribution, energy spectrum and the fractional contributions of each individual CM. A multiple-source model has been further developed to reconstruct the original phase space. Separate sources were created with accurate source intensity, energy, fluence and angular distributions for the target, primary collimator and flattening filter. Good agreement (within 2%) between the Monte Carlo calculations with the source model and those with the original phase space was achieved in the dose distributions for field sizes of 4 cm x 4 cm to 40 cm x 40 cm at source surface distances (SSDs) of 80-120 cm. The dose distributions in lung and bone heterogeneous phantoms have also been found to be in good agreement (within 2%) for 4, 6 and 15 MV photon beams for various field sizes between the Monte Carlo calculations with the source model and those with the original phase space.


Assuntos
Fótons/uso terapêutico , Planejamento da Radioterapia Assistida por Computador , Modelos Teóricos , Método de Monte Carlo , Aceleradores de Partículas , Imagens de Fantasmas , Software
14.
Phys Med Biol ; 46(4): 1039-60, 2001 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-11324950

RESUMO

We have investigated the tongue-and-groove effect on the IMRT dose distributions for a Varian MLC. We have compared the dose distributions calculated using the intensity maps with and without the tongue-and-groove effect. Our results showed that, for one intensity-modulated treatment field, the maximum tongue-and-groove effect could be up to 10% of the maximum dose in the dose distributions. For an IMRT treatment with multiple gantry angles (> or = 5), the difference between the dose distributions with and without the tongue-and-groove effect was hardly visible, less than 1.6% for the two typical clinical cases studied. After considering the patient setup errors, the dose distributions were smoothed with reduced and insignificant differences between plans with and without the tongue-and-groove effect. Therefore, for a multiple-field IMRT plan (> or = 5), the tongue-and-groove effect on the IMRT dose distributions will be generally clinically insignificant due to the smearing effect of individual fields. The tongue-and-groove effect on an IMRT plan with small number of fields (< 5) will vary depending on the number of fields in a plan (coplanar or non-coplanar), the MLC leaf sequences and the patient setup uncertainty, and may be significant (> 5% of maximum dose) in some cases, especially when the patient setup uncertainty is small (< or = 2 mm).


Assuntos
Radioterapia Conformacional/instrumentação , Radioterapia Conformacional/métodos , Neoplasias Ósseas/radioterapia , Elétrons , Humanos , Masculino , Método de Monte Carlo , Fótons , Neoplasias da Próstata/radioterapia , Radiometria/métodos , Planejamento da Radioterapia Assistida por Computador
15.
Phys Med Biol ; 46(5): 1429-49, 2001 May.
Artigo em Inglês | MEDLINE | ID: mdl-11384063

RESUMO

A method for deriving the electron and photon energy spectra from electron beam central axis percentage depth dose (PDD) curves has been investigated. The PDD curves of 6, 12 and 20 MeV electron beams obtained from the Monte Carlo full phase space simulations of the Varian linear accelerator treatment head have been used to test the method. We have employed a 'random creep' algorithm to determine the energy spectra of electrons and photons in a clinical electron beam. The fitted electron and photon energy spectra have been compared with the corresponding spectra obtained from the Monte Carlo full phase space simulations. Our fitted energy spectra are in good agreement with the Monte Carlo simulated spectra in terms of peak location, peak width, amplitude and smoothness of the spectrum. In addition, the derived depth dose curves of head-generated photons agree well in both shape and amplitude with those calculated using the full phase space data. The central axis depth dose curves and dose profiles at various depths have been compared using an automated electron beam commissioning procedure. The comparison has demonstrated that our method is capable of deriving the energy spectra for the Varian accelerator electron beams investigated. We have implemented this method in the electron beam commissioning procedure for Monte Carlo electron beam dose calculations.


Assuntos
Elétrons , Fótons , Planejamento da Radioterapia Assistida por Computador , Algoritmos , Processamento de Imagem Assistida por Computador , Análise dos Mínimos Quadrados , Método de Monte Carlo
16.
Phys Med Biol ; 45(9): 2483-95, 2000 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-11008950

RESUMO

The purpose of this work was to use Monte Carlo simulations to verify the accuracy of the dose distributions from a commercial treatment planning optimization system (Corvus, Nomos Corp., Sewickley, PA) for intensity-modulated radiotherapy (IMRT). A Monte Carlo treatment planning system has been implemented clinically to improve and verify the accuracy of radiotherapy dose calculations. Further modifications to the system were made to compute the dose in a patient for multiple fixed-gantry IMRT fields. The dose distributions in the experimental phantoms and in the patients were calculated and used to verify the optimized treatment plans generated by the Corvus system. The Monte Carlo calculated IMRT dose distributions agreed with the measurements to within 2% of the maximum dose for all the beam energies and field sizes for both the homogeneous and heterogeneous phantoms. The dose distributions predicted by the Corvus system, which employs a finite-size pencil beam (FSPB) algorithm, agreed with the Monte Carlo simulations and measurements to within 4% in a cylindrical water phantom with various hypothetical target shapes. Discrepancies of more than 5% (relative to the prescribed target dose) in the target region and over 20% in the critical structures were found in some IMRT patient calculations. The FSPB algorithm as implemented in the Corvus system is adequate for homogeneous phantoms (such as prostate) but may result in significant under or over-estimation of the dose in some cases involving heterogeneities such as the air-tissue, lung-tissue and tissue-bone interfaces.


Assuntos
Simulação por Computador , Planejamento da Radioterapia Assistida por Computador , Radioterapia Conformacional/métodos , Humanos , Processamento de Imagem Assistida por Computador , Método de Monte Carlo , Imagens de Fantasmas , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/normas , Radioterapia Conformacional/normas , Reprodutibilidade dos Testes , Água
17.
Phys Med Biol ; 45(8): 2293-311, 2000 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-10958195

RESUMO

This work investigates the feasibility of optimizing energy- and intensity-modulated electron beams for radiation therapy. A multileaf collimator (MLC) specially designed for modulated electron radiotherapy (MERT) was investigated both experimentally and by Monte Carlo simulations. An inverse-planning system based on Monte Carlo dose calculations was developed to optimize electron beam energy and intensity to achieve dose conformity for target volumes near the surface. The results showed that an MLC with 5 mm leaf widths could produce complex field shapes for MERT. Electron intra- and inter-leaf leakage had negligible effects on the dose distributions delivered with the MLC, even at shallow depths. Focused leaf ends reduced the electron scattering contributions to the dose compared with straight leaf ends. As anticipated, moving the MLC position toward the patient surface reduced the penumbra significantly. There were significant differences in the beamlet distributions calculated by an analytic 3-D pencil beam algorithm and the Monte Carlo method. The Monte Carlo calculated beamlet distributions were essential to the accuracy of the MERT dose distribution in cases involving large air gaps, oblique incidence and heterogeneous treatment targets (at the tissue-bone and bone-lung interfaces). To demonstrate the potential of MERT for target dose coverage and normal tissue sparing for treatment of superficial targets, treatment plans for a hypothetical treatment were compared using photon beams and MERT.


Assuntos
Elétrons/uso terapêutico , Radioterapia Conformacional/instrumentação , Radioterapia Conformacional/métodos , Algoritmos , Neoplasias da Mama/diagnóstico por imagem , Neoplasias da Mama/radioterapia , Simulação por Computador , Feminino , Humanos , Neoplasias Pulmonares/diagnóstico por imagem , Neoplasias Pulmonares/radioterapia , Método de Monte Carlo , Imagens de Fantasmas , Fótons/uso terapêutico , Radiografia , Radiometria , Espalhamento de Radiação
18.
Phys Med Biol ; 48(5): 561-72, 2003 Mar 07.
Artigo em Inglês | MEDLINE | ID: mdl-12696795

RESUMO

This paper investigates a quality assurance (QA) phantom specially designed to verify the accuracy of dose distributions and monitor units (MU) calculated by clinical treatment planning optimization systems and by the Monte Carlo method for intensity-modulated radiotherapy (IMRT). The QA phantom is a PMMA cylinder of 30 cm diameter and 40 cm length with various bone and lung inserts. A procedure (and formalism) has been developed to measure the absolute dose to water in the PMMA phantom. Another cylindrical phantom of the same dimensions, but made of water, was used to confirm the results obtained with the PMMA phantom. The PMMA phantom was irradiated by 4, 6 and 15 MV photon beams and the dose was measured using an ionization chamber and compared to the results calculated by a commercial inverse planning system (CORVUS, NOMOS, Sewickley, PA) and by the Monte Carlo method. The results show that the dose distributions calculated by both CORVUS and Monte Carlo agreed to within 2% of dose maximum with measured results in the uniform PMMA phantom for both open and intensity-modulated fields. Similar agreement was obtained between Monte Carlo calculations and measured results with the bone and lung heterogeneity inside the PMMA phantom while the CORVUS results were 4% different. The QA phantom has been integrated as a routine QA procedure for the patient's IMRT dose verification at Stanford since 1999.


Assuntos
Imagens de Fantasmas/normas , Garantia da Qualidade dos Cuidados de Saúde/normas , Radiometria/normas , Planejamento da Radioterapia Assistida por Computador/normas , Radioterapia Conformacional/instrumentação , Radioterapia Conformacional/normas , Desenho de Equipamento , Polimetil Metacrilato , Garantia da Qualidade dos Cuidados de Saúde/métodos , Controle de Qualidade , Radiometria/instrumentação , Radiometria/métodos , Dosagem Radioterapêutica/normas , Planejamento da Radioterapia Assistida por Computador/instrumentação , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia Conformacional/métodos , Padrões de Referência , Estados Unidos
19.
Phys Med Biol ; 47(10): 1671-89, 2002 May 21.
Artigo em Inglês | MEDLINE | ID: mdl-12069086

RESUMO

A Monte Carlo user code, MCDOSE, has been developed for radiotherapy treatment planning (RTP) dose calculations. MCDOSE is designed as a dose calculation module suitable for adaptation to host RTP systems. MCDOSE can be used for both conventional photon/electron beam calculation and intensity modulated radiotherapy (IMRT) treatment planning. MCDOSE uses a multiple-source model to reconstruct the treatment beam phase space. Based on Monte Carlo simulated or measured beam data acquired during commissioning, source-model parameters are adjusted through an automated procedure. Beam modifiers such as jaws, physical and dynamic wedges, compensators, blocks, electron cut-outs and bolus are simulated by MCDOSE together with a 3D rectilinear patient geometry model built from CT data. Dose distributions calculated using MCDOSE agreed well with those calculated by the EGS4/DOSXYZ code using different beam set-ups and beam modifiers. Heterogeneity correction factors for layered-lung or layered-bone phantoms as calculated by both codes were consistent with measured data to within 1%. The effect of energy cut-offs for particle transport was investigated. Variance reduction techniques were implemented in MCDOSE to achieve a speedup factor of 10-30 compared to DOSXYZ.


Assuntos
Método de Monte Carlo , Neoplasias/radioterapia , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia Conformacional/métodos , Simulação por Computador , Humanos , Imagens de Fantasmas
20.
Med Dosim ; 26(2): 157-68, 2001.
Artigo em Inglês | MEDLINE | ID: mdl-11444518

RESUMO

This article describes photon beam Monte Carlo simulation for multi leaf collimator (MLC)-based intensity-modulated radiotherapy (IMRT). We present the general aspects of the Monte Carlo method for the non-Monte Carloist with an emphasis given to patient-specific radiotherapy application. Patient-specific application of the Monte Carlo method can be used for IMRT dose verification, inverse planning, and forward planning in conventional conformal radiotherapy. Because it is difficult to measure IMRT dose distributions in heterogeneous phantoms that approximate a patient, Monte Carlo methods can be used to verify IMRT dose distributions that are calculated using conventional methods. Furthermore, using Monte Carlo as the dose calculation method for inverse planning results in better-optimized treatment plans. We describe both aspects and present our recent results to illustrate the discussion. Finally, we present current issues related to clinical implementation of Monte Carlo dose calculation. Monte Carlo is the most recent, and most accurate, method of radiotherapy dose calculation. It is currently in the process of being implemented by various treatment planning vendors and will be available for clinical use in the immediate future.


Assuntos
Método de Monte Carlo , Radioterapia Conformacional/estatística & dados numéricos , Humanos , Tomografia Computadorizada por Raios X