1.
Opt Lett
; 46(2): 424-427, 2021 Jan 15.
Artigo
em Inglês
| MEDLINE
| ID: mdl-33449045
RESUMO
Fast time-gated single-photon detectors demonstrated high depth sensitivity in the detection of localized absorption perturbations inside scattering media, but their use for in vivo clinical applications-such as functional imaging of brain activation-was impaired by their small (<0.04mm2) active area. Here, we demonstrate, both on phantoms and in vivo, the performance of a fast-gated digital silicon photomultiplier (SiPM) that features an overall active area of 8.6mm2, overcoming the photon collection capability of established time-gated single-pixel detectors by orders of magnitude, enabling deep investigations within scattering media and high signal-to-noise ratios at late photon arrival times.