Your browser doesn't support javascript.
loading
Long-term clinical experience with the Omnicarbon prosthetic valve.
Torregrosa, S; Gómez-Plana, J; Valera, F J; Caffarena, J; Maroñas, J M; García-Sánchez, F; Peris, J; Frías, R; Caffarena, J M.
Afiliação
  • Torregrosa S; Servicio de Cirugía Cardiovascular, Hospital Universitario La Fé, Valencia, Spain.
Ann Thorac Surg ; 68(3): 881-6, 1999 Sep.
Article em En | MEDLINE | ID: mdl-10509978
ABSTRACT

BACKGROUND:

From February 1985 to December 1994, 781 Omnicarbon valve prostheses were implanted in 647 patients. These were 357 male and 290 female patients with a mean age of 53.5+/-10.5 years (range, 4 to 78 years). Before operation, 81% of the patients were in New York Heart Association class III or IV, 16% were in class II, and only 3% were in class I.

METHODS:

There were 227 aortic valve replacements (AVR) (35%), 286 mitral valve replacements (MVR) (44%), and 134 double-valve replacements (DVR) (21%) (AVR + MVR). Follow-up was 96.3% complete and consisted of 2,746 patient-years (mean follow-up, 4.6 years, and maximum follow-up, 10.7 years).

RESULTS:

Hospital mortality rates were 7.0% for AVR, 8.0% for MVR, and 8.2% for DVR. The annualized rate of anticoagulant-related hemorrhage was 0.8% per patient-year, and thromboembolism occurred at a rate of 0.7% per patient-year. No structural failure was observed during 10-year follow-up. Twenty-one instances of nonstructural dysfunction (two, pannus growth, and 19, dehiscence) of the Omnicarbon valve occurred in 20 patients, an incidence of 0.8% per patient-year. Hemolytic anemia was observed only in the presence of valvular dehiscence (6 of 19). Eight patients (0.3% per patient-year) had development of prosthetic valve endocarditis (4, AVR; 2, MVR; and 2 DVR). At the end of 10 years of follow-up, 91% of the survivors were in New York Heart Association class I or II. The overall survival rate at 10 years was 82.5%+/-2.6% (85.0%+/-3.9%, AVR; 81.0%+/-4.1%, MVR; and 82.5%+/-2.6%, DVR). Considering only valve-related deaths, the survival rate at 10 years was 91.9%+/-2.4% (90.0%+/-2.7%, AVR; 93.1%+/-3.8%, MVR; and 90.0%+/-1.8%, DVR).

CONCLUSIONS:

Clinical results over a 10-year follow-up are excellent with the Omnicarbon prosthesis.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Male / Middle aged Idioma: En Revista: Ann Thorac Surg Ano de publicação: 1999 Tipo de documento: Article País de afiliação: Espanha
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Male / Middle aged Idioma: En Revista: Ann Thorac Surg Ano de publicação: 1999 Tipo de documento: Article País de afiliação: Espanha