Three body wear studies of five dental composites preconditioned in food simulating media.
Mair, L H; Krishnan, V K.
Afiliação
  • Mair LH; Department of Clinical Engineering, University of Liverpool, UK.
Biomed Mater Eng ; 9(3): 145-9, 1999.
Article em En | MEDLINE | ID: mdl-10572618
ABSTRACT
Progressive three body wear up to 30 min for five dental composites (Concise, Heliomolar, Occlusin, P-30, Profile, TPH) stored for 9 months at 37 degrees C in citrate, lactate and PBS buffers, 25 and 75% ethanol media, has been measured using a wear tester. Particle size of the filler and the storage media are found to play dominant roles in determining the extent of wear. Heliomolar, which is a microfilled composite, shows maximum wear whereas wear is least in composites containing filler particles of conventional and hybrid size such as P-30 (> 1 micron). The storage media is found to attack and degrade glass filler particles compared to composites having quartz fillers.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Resinas Compostas / Análise do Estresse Dentário Tipo de estudo: Prognostic_studies Idioma: En Revista: Biomed Mater Eng Assunto da revista: BIOTECNOLOGIA / ENGENHARIA BIOMEDICA Ano de publicação: 1999 Tipo de documento: Article País de afiliação: Reino Unido
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Resinas Compostas / Análise do Estresse Dentário Tipo de estudo: Prognostic_studies Idioma: En Revista: Biomed Mater Eng Assunto da revista: BIOTECNOLOGIA / ENGENHARIA BIOMEDICA Ano de publicação: 1999 Tipo de documento: Article País de afiliação: Reino Unido