ABSTRACT
The wide availability of fibrinogen estimations based on the prothrombin time (PT-Fg) has caused concern about the variability and clinical utility of fibrinogen assays. In a multi-centre study, we investigated fibrinogen assays using various reagents and analysers. Clauss assays generally gave good agreement, although one reagent gave 15-30% higher values in DIC and thrombolysis. Two commercial reference preparations had much lower potencies than the manufacturers declared, and plasma turbidity influenced parallelism in some Clauss assays. PT-Fg assays gave higher values than Clauss and showed calibrant dependent effects, the degree of disparity correlating with calibrant and test sample turbidity. Analyser and thromboplastin dependent differences were noted. The relationship between Clauss and PT-Fg assays was sigmoid, and the plateau of maximal PT-Fg differed by about 2 g/l between reagents. ELISA and immunonephelometric assays correlated well, but with a high degree of scatter. Antigen levels were higher than Clauss, but slightly lower than PT-Fg assays, which appeared to be influenced by degraded fibrinogen. Clauss assays are generally reproducible between centres, analysers and reagents, but PT-Fg assays are not reliable in clinical settings.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Kit de Reagentes para Diagnóstico / Fibrinogênio Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Thromb Haemost Ano de publicação: 2002 Tipo de documento: Article País de afiliação: Reino Unido
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Kit de Reagentes para Diagnóstico / Fibrinogênio Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Thromb Haemost Ano de publicação: 2002 Tipo de documento: Article País de afiliação: Reino Unido