Your browser doesn't support javascript.
loading
Range, range modulation, and field radius requirements for proton therapy of prostate cancer.
Moyers, M F; Miller, D W.
Afiliação
  • Moyers MF; Loma Linda University Medical Center, Loma Linda, California 92354, USA. moyers@proton.llumc.edu
Technol Cancer Res Treat ; 2(5): 445-7, 2003 Oct.
Article em En | MEDLINE | ID: mdl-14529309
ABSTRACT
The Loma Linda University Proton Treatment Facility has treated over 5,000 patients for prostate cancer. Other institutions may find information regarding field size and range requirements for this population of patients useful for designing new proton beamlines. The maximum range, range modulation, and maximum field radius for 240 fields of prostate patients undergoing treatment were sampled and analyzed. Most fields required a range less than 290 mm of water, a modulation width less than or equal to 120 mm, and a radius less than 75 mm.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Próstata / Radioterapia (Especialidade) / Terapia com Prótons Limite: Humans / Male Idioma: En Revista: Technol Cancer Res Treat Assunto da revista: NEOPLASIAS / TERAPEUTICA Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias da Próstata / Radioterapia (Especialidade) / Terapia com Prótons Limite: Humans / Male Idioma: En Revista: Technol Cancer Res Treat Assunto da revista: NEOPLASIAS / TERAPEUTICA Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Estados Unidos