Your browser doesn't support javascript.
loading
[Phantom study for the detection of simulated lesions in five different digital and one conventional mammography system]. / Phantomstudie zur Detektion simulierter Läsionen an fünf verschiedenen digitalen und einem konventionellen Mammographiesystem.
Schulz-Wendtland, R; Hermann, K-P; Lell, M; Böhner, C; Wenkel, E; Imhoff, K; Schmid, A; Krug, B; Bautz, W.
Afiliação
  • Schulz-Wendtland R; Radiologisches Institut, Universität Erlangen-Nürnberg. ruediger.schulz-wendtland@idr.imed.uni-erlangen.de
Rofo ; 176(8): 1127-32, 2004 Aug.
Article em De | MEDLINE | ID: mdl-15346289
ABSTRACT

PURPOSE:

Experimental phantom study for the detection of simulated lesions with five different digital and one conventional screen-film mammography system. MATERIALS AND

METHODS:

Three radiographs were obtained at various configurations of the phantom with one conventional screen-film system (Mammomat 3000 N) (Siemens), five digital systems (high resolution computed radiography system [Fuji/Siemens], one a-Si detector [GE Medical Systems], two a-Se detectors [Siemens; Hologic / Lorad] and one CCD detector [Fischer Imaging]), applying the same exposure parameters. The Wisconsin Mammographic Random Phantom, Model 152 A, was used. Five investigators with different experience in mammography (3 months to more than 4 years) evaluated the 18 randomly selected radiographs.

RESULTS:

No significant differences were found in the detection rate of simulated breast lesions for conventional screen-film mammography (84.9 %), high resolution computed radiography (86.7 %) and digital mammography with an a-Si detector (89.8 %). Highly significantly better results (p < 0.001) were found with the two a-Se systems (97.3 %) and the CCD system (100 %).

CONCLUSION:

Conventional screen-film mammography can be replaced by high resolution computed radiography and digital mammography with a-Si-, a-Se- and CCD-detectors. This has to be confirmed in further clinical studies.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Mamografia / Xeromamografia / Imagens de Fantasmas Tipo de estudo: Diagnostic_studies Limite: Female / Humans Idioma: De Revista: Rofo Ano de publicação: 2004 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Mamografia / Xeromamografia / Imagens de Fantasmas Tipo de estudo: Diagnostic_studies Limite: Female / Humans Idioma: De Revista: Rofo Ano de publicação: 2004 Tipo de documento: Article