[Digital mammography with high-resolution storage plates (CR) versus full-field digital mammography (CCD) (DR) for microcalcifications and focal lesions -- a retrospective clinical histologic analysis (n = 102)]. / Digitale Mammographie mit hochauflösenden Speicherfolien (CR) versus digitaler Vollfeldmammographie (CCD) (DR): Mikrokalzifikationen und Herdbefunde -- eine retrospektiv klinisch-histologische Analyse (n = 102).
Schulz-Wendtland, R; Lell, M; Wenkel, E; Böhner, C; Dassel, M-S; Bautz, W.
Afiliação
  • Schulz-Wendtland R; Radiologisches Institut, Universität Erlangen-Nürnberg. ruediger.schulz-wendtland@idr.imed.uni-erlangen.de
Rofo ; 177(1): 67-71, 2005 Jan.
Article em De | MEDLINE | ID: mdl-15657822
ABSTRACT

PURPOSE:

To determine the diagnostic accuracy of microcalcifications and focal lesions in a retrospective clinical-histological study using high-resolution digital phosphor storage plates (hard copy) and full-field digital mammography (hard copy). MATERIALS AND

METHODS:

From May 2003 to September 2003, 102 patients underwent digital storage plate mammography (CR), using a mammography unit (Mammomat 3000 N, Siemens) in combination with a high resolution (9 lp/mm) digital storage phosphor plate system (pixel size 50 microm) (Fuji/Siemens). After diagnosis and preoperative wire localization, full-field digital mammography (CCD) (DR) was performed with the same exposure parameters. The full-field digital mammography used a CCD-detector (SenoScan) (Fisher Imaging) with a resolution of 10 Ip/mm and a pixel size of 50 microm. Five investigators determined the diagnosis (BI-RADS I - V) retrospectively after the operation from randomly distributed mediolateral views (hard copy reading). These results were correlated with the final histology.

RESULTS:

The diagnostic accuracy of digital storage plate mammography (CR) and full-field digital mammography (CCD) (DR) was 73 % and 71 % for all findings (n = 102), 73 % and 71 % for microcalcifications (n = 51), and 72 % and 70 % for focal lesions (n = 51). The overall results showed no difference.

CONCLUSION:

Our findings indicate the equivalence of high-resolution digital phosphor storage plate mammography (CR) and full-field digital mammography (CCD) (DR).
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doenças Mamárias / Neoplasias da Mama / Calcinose / Mamografia / Intensificação de Imagem Radiográfica Tipo de estudo: Diagnostic_studies / Evaluation_studies / Observational_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: De Revista: Rofo Ano de publicação: 2005 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doenças Mamárias / Neoplasias da Mama / Calcinose / Mamografia / Intensificação de Imagem Radiográfica Tipo de estudo: Diagnostic_studies / Evaluation_studies / Observational_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: De Revista: Rofo Ano de publicação: 2005 Tipo de documento: Article