[Controlling of the high-voltage generator in a direct radiography system].
Zhongguo Yi Liao Qi Xie Za Zhi
; 30(2): 111-3, 2006 Mar.
Article
em Zh
| MEDLINE
| ID: mdl-16830803
ABSTRACT
A proposal is presented to improve the original complicated control of the exposure parameters in the direct radiography system (DR). The interface circuitry of the high-voltage generator is redesigned using RS-232 serial COM port. The data and signals are transmitted between the personal computer and the high-voltage generator in the serial mode. A program package is designed to realize the control of the exposure parameters of the high-voltage generator in DR on the PC machine. Experiment results have shown that the proposed console operates steadily and it is capable of providing a more convenient operation console for the direct radiography system.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Software
/
Radiografia
/
Interpretação de Imagem Radiográfica Assistida por Computador
Tipo de estudo:
Diagnostic_studies
Limite:
Humans
Idioma:
Zh
Revista:
Zhongguo Yi Liao Qi Xie Za Zhi
Ano de publicação:
2006
Tipo de documento:
Article