Comparison of two portable solid state detectors with an improved collimation and alignment device for mammographic x-ray spectroscopy.
Bottigli, U; Golosio, B; Masala, G L; Oliva, P; Stumbo, S; Delogu, P; Fantacci, M E; Abbene, L; Fauci, F; Raso, G.
Afiliação
  • Bottigli U; Struttura Dipartimentale di Matematica e Fisica dell'Università degli Studi di Sassari and Sezione I.N.F.N., Cagliari, Italy.
Med Phys ; 33(9): 3469-77, 2006 Sep.
Article em En | MEDLINE | ID: mdl-17022243
ABSTRACT
We describe a portable system for mammographic x-ray spectroscopy, based on a 2 X 2 X 1 mm3 cadmium telluride (CdTe) solid state detector, that is greatly improved over a similar system based on a 3 X 3 X 2 mm3 cadmium zinc telluride (CZT) solid state detector evaluated in an earlier work. The CdTe system utilized new pinhole collimators and an alignment device that facilitated measurement of mammographic x-ray spectra. Mammographic x-ray spectra acquired by each system were comparable. Half value layer measurements obtained using an ion chamber agreed closely with those derived from the x-ray spectra measured by either detector. The faster electronics and other features of the CdTe detector allowed its use with a larger pinhole collimator than could be used with the CZT detector. Additionally, the improved pinhole collimator and alignment features of the apparatus permitted much more rapid setup for acquisition of x-ray spectra than was possible on the system described in the earlier work. These improvements in detector technology, collimation and ease of alignment, as well as low cost, make this apparatus attractive as a tool for both laboratory research and advanced mammography quality control.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Espectrometria por Raios X / Mamografia / Análise de Falha de Equipamento Tipo de estudo: Diagnostic_studies Idioma: En Revista: Med Phys Ano de publicação: 2006 Tipo de documento: Article País de afiliação: Itália
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Espectrometria por Raios X / Mamografia / Análise de Falha de Equipamento Tipo de estudo: Diagnostic_studies Idioma: En Revista: Med Phys Ano de publicação: 2006 Tipo de documento: Article País de afiliação: Itália