[Investigation on baseline correction in pollutant monitoring on-line by NIR spectroscopy].
Sun, Yi; Du, Zhen-hui; Yin, Xin; Xu, Ke-xin.
Afiliação
  • Sun Y; State Key Lab of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China. sunyi820710@hotmail.com
Guang Pu Xue Yu Guang Pu Fen Xi ; 28(10): 2282-4, 2008 Oct.
Article em Zh | MEDLINE | ID: mdl-19123389
ABSTRACT
Spectroscopic data pre-processing is important to spectra scaling and quantitative analysis, and correction of baseline drift caused by variation of measuring condition is critical in NIR spectroscopy analysis in-situ. In the present paper, the NIR absorption spectra of propane and isobutane with different concentration were detected by BRUKER EQUINOX55 spectrometer. The "Derivative-smooth" method and the baseline correction method which is accessory of the instrument were theoretically analyzed, and the data processed by the two kinds of method were compared. The results show that the method that is accessory of the instrument can correct steady-state and linear drift, and can gain high SNR, high accuracy; "Derivative-smooth" method has better applicability, can correct non-steady-state and nonlinear drift better, and is a good kinds of pre-processing method in analysis in situ.
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Idioma: Zh Revista: Guang Pu Xue Yu Guang Pu Fen Xi Ano de publicação: 2008 Tipo de documento: Article País de afiliação: China
Buscar no Google