A study of experimental measurements of dosimetric parameters in HDR IR-192 source.
Med Dosim
; 35(4): 250-4, 2010.
Article
em En
| MEDLINE
| ID: mdl-19944591
ABSTRACT
Thermoluminescent dosimeters have been used to perform dosimetry measurements for the widely used Ir-192 microSelectron-HDR source with an improved polystyrene phantom. Radial dose functions and anisotropy functions, main parameters of 2D dose-rate formalism from the TG-43U1 protocol, have been obtained experimentally. Measurement results are compared with that of the Monte Carlo calculations reported, and no difference has been found between them.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Dosimetria Termoluminescente
/
Radioisótopos de Irídio
Tipo de estudo:
Diagnostic_studies
Idioma:
En
Revista:
Med Dosim
Assunto da revista:
RADIOTERAPIA
Ano de publicação:
2010
Tipo de documento:
Article
País de afiliação:
China