Your browser doesn't support javascript.
loading
Intercomparison of methods for coincidence summing corrections in gamma-ray spectrometry.
Lépy, M-C; Altzitzoglou, T; Anagnostakis, M J; Arnold, D; Capogni, M; Ceccatelli, A; De Felice, P; Dersch, R; Dryak, P; Fazio, A; Ferreux, L; Guardati, M; Han, J B; Hurtado, S; Karfopoulos, K L; Klemola, S; Kovar, P; Lee, K B; Ocone, R; Ott, O; Sima, O; Sudar, S; Svec, A; Tao, Chau Van; Thanh, Tran Thien; Vidmar, T.
Afiliação
  • Lépy MC; Laboratoire National Henri Becquerel (LNE-LNHB), CEA, LIST, Gif-sur-Yvette, France. marie-christine.lepy@cea.fr
Appl Radiat Isot ; 68(7-8): 1407-12; discussion 1412, 2010.
Article em En | MEDLINE | ID: mdl-20117939
ABSTRACT
A comparison of the coincidence summing correction methods is presented. Since there are several ways for computing these corrections, each method has advantages and drawbacks that could be compared. This part of the comparison was restricted to point sources. The same experimental spectra, decay scheme and photon emission intensities were used by all the participants. The results were expressed as coincidence summing correction factors for several energies of (152)Eu and (134)Cs, and three source-to-detector distances. They are presented and discussed.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Appl Radiat Isot Assunto da revista: MEDICINA NUCLEAR / SAUDE AMBIENTAL Ano de publicação: 2010 Tipo de documento: Article País de afiliação: França