Your browser doesn't support javascript.
loading
Evaluation of a patch antenna applicator for time reversal hyperthemia.
Dobsícek Trefná, Hana; Vrba, Jan; Persson, Mikael.
Afiliação
  • Dobsícek Trefná H; Department of Signals and Systems, Chalmers University of Technology, Gothenburg, Sweden. hanatre@chalmers.se
Int J Hyperthermia ; 26(2): 185-97, 2010.
Article em En | MEDLINE | ID: mdl-20146572
ABSTRACT

PURPOSE:

To describe the design, analysis and evaluation of a new antenna array system for microwave hyperthermia. The proposed antenna array was evaluated by the focusing method based on the time-reversal principle. MATERIALS AND

METHODS:

Power absorption distributions in a cylindrical homogeneous and inhomogeneous phantom were calculated for the frequency range 500-900 MHz. Two set-ups with 12 and 16 antennas were analysed by comparing the changes in focusing areas enclosed by the 50%, 75% and 90% iso-SAR contours. For a more quantitative evaluation of the results the average power absorption ratio and remaining tissue maximum index were calculated.

RESULTS:

The sharpest focusing area in the centre of the phantom, 151 mm(2) (9 x 20 mm) (90% iso-SAR), was obtained by using 16 antennas at frequency 900 MHz. The largest focusing area of 280 mm(2) (13 x 24 mm) (90% iso-SAR) was obtained by using 16 antennas at 500 MHz. The SAR focus was steered in the desired radial direction obtaining a 43 mm(2) 90% iso-SAR focus-width in a semi-three-dimensional neck phantom. The results showed qualitative agreement between three dimensions (3D) and two dimensions (2D) for the performance indicators.

CONCLUSIONS:

The conducted study confirms the feasibility of the time-reversal-based focusing methods for microwave hyperthermia. The proposed system shows promise and is suitable for further development in the treatment of head and neck tumours, and extremities application.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imagens de Fantasmas / Desenho de Equipamento / Hipertermia Induzida Tipo de estudo: Evaluation_studies / Qualitative_research Limite: Humans Idioma: En Revista: Int J Hyperthermia Assunto da revista: NEOPLASIAS / TERAPEUTICA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Suécia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imagens de Fantasmas / Desenho de Equipamento / Hipertermia Induzida Tipo de estudo: Evaluation_studies / Qualitative_research Limite: Humans Idioma: En Revista: Int J Hyperthermia Assunto da revista: NEOPLASIAS / TERAPEUTICA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Suécia