X-ray streak cameras are extensively used to study transient x-ray processes in experiments carried out on various laser facilities at the Research Center of Laser Fusion. Precise calibration and clear description of the sweep speed of the x-ray streak camera are vital for obtaining precise temporal information to understand the fast physics phenomena. An x-ray streak camera named wide-slit x-ray streak camera has been calibrated using a small-scale laser facility with pulse duration of 8 ps. Sweep speeds on most part of the output screen (charge coupled device) are obtained. These calibration results can be used to correct the nonlinearity of sweep speed in measurement of temporal processes.