Clinical effect of implantable cardioverter defibrillator replacements: when should you resume driving after an implantable cardioverter defibrillator replacement?
Kawata, Hiro; Noda, Takashi; Kurita, Takashi; Yamagata, Kenichiro; Yamada, Yuko; Okamura, Hideo; Satomi, Kazuhiro; Shimizu, Wataru; Suyama, Kazuhiro; Aihara, Naohiko; Isobe, Mitsuaki; Kamakura, Shiro.
Afiliação
  • Kawata H; Division of Arrhythmia and Electrophysiology, Department of Cardiovascular Medicine, National Cerebral and Cardiovascular Center, Tokyo Medical and Dental University, Tokyo, Japan.
Circ J ; 74(11): 2301-7, 2010 Nov.
Article em En | MEDLINE | ID: mdl-20877125
ABSTRACT

BACKGROUND:

The intervals of the driving restrictions after an implantable cardioverter defibrillator (ICD) replacement vary across the different countries around the world. However, little is known regarding the appropriate duration for driving restrictions after an ICD replacement. The aim of this study was to investigate the clinical effect of ICD replacements and to elucidate when to resume driving an automobile after an ICD replacement. METHODS AND

RESULTS:

The study reviewed 139 consecutive patients with an ICD replacement in order to evaluate the incidence of ICD therapies before and after ICD replacements, and to assess the time-dependence of the ICD therapies after the ICD replacement. There was no significant difference in the incidence of ICD therapies delivered during durations of 3 months and 6 months before and after the ICD replacement (P=0.28, and 1.0, respectively). ICD therapies after the replacements were observed in 8.6% of the patients who were legally eligible to drive according to the Japanese guidelines at 1 year, and that was associated with a relatively low annual risk of death or injury to others.

CONCLUSIONS:

Implantable cardioverter defibrillator replacements did not affect the future ICD therapies under similar algorithms. The appropriate interval for driving restrictions after an ICD replacement is recommended to be a week or so, with a system integrity check performed before resumption of driving.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Condução de Veículo / Cardioversão Elétrica / Acidentes de Trânsito / Desfibriladores Implantáveis Tipo de estudo: Etiology_studies / Guideline / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged País/Região como assunto: Asia Idioma: En Revista: Circ J Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Japão
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Condução de Veículo / Cardioversão Elétrica / Acidentes de Trânsito / Desfibriladores Implantáveis Tipo de estudo: Etiology_studies / Guideline / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged País/Região como assunto: Asia Idioma: En Revista: Circ J Assunto da revista: ANGIOLOGIA / CARDIOLOGIA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Japão