Your browser doesn't support javascript.
loading
Performance of new enzyme-linked fluorescent assays for detection of Epstein-Barr virus specific antibodies in routine diagnostics.
Koidl, Christoph; Riedl, Regina; Schweighofer, Bettina; Fett, Sandra; Bozic, Michael; Marth, Egon.
Afiliação
  • Koidl C; Department of Virology and Serological Diagnostics, Institute of Hygiene, Microbiology and Environmental Medicine, Medical University of Graz, Austria. christoph.koidl@medunigraz.at
Wien Klin Wochenschr ; 123(7-8): 230-4, 2011 Apr.
Article em En | MEDLINE | ID: mdl-21451951
ABSTRACT
BACKGROUND AND

AIMS:

Indirect fluorescent antibody assays are still the gold standard for the detection of Epstein-Barr-virus-specific antibodies; however, this technique requires several manual steps and an experienced technician. This retrospective study investigated the performance of the new VIDAS(®) enzyme-linked fluorescent assays for automated qualitative detection of EBV VCA IgM, VCA/EA IgG, and EBNA IgG antibodies in routine diagnostics.

METHODS:

174 serum samples were tested first with the gold standard. In context with the clinical status, 60 samples IFA IgG/IgM positive and with clinical symptoms compatible with infectious mononucleosis, 26 samples IFA IgG/IgM negative and missing any clinical symptoms and 88 samples with varying IFA status and without any clinical information were defined. In a second step all samples were retested with the new assays.

RESULTS:

In the overall agreement between VIDAS(®) and IFA for evaluable results, almost perfect agreement was observed (kappa = 0.91; 95% confidence interval (CI), 0.86-0.97). Estimating all indeterminate VIDAS(®) results as discordant or concordant the observed kappa values were 0.71 (CI 0.63-0.79) and 0.93 (CI 0.88-0.98), respectively. With the new assays 45, 22, and 70 identical results were obtained, respectively. Western blot analysis of the discrepant samples showed a quasi similar performance of both assays.

CONCLUSIONS:

The new VIDAS(®) assays can be an alternative to IFA testing especially in high-throughput laboratories. Full automation of EBV serological diagnostis by the new VIDAS assays is of major importance for routine diagnostic laboratories.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Fluorescência / Ensaio de Imunoadsorção Enzimática / Herpesvirus Humano 4 / Infecções por Vírus Epstein-Barr / Testes Diagnósticos de Rotina / Anticorpos Antivirais Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies / Qualitative_research Limite: Humans Idioma: En Revista: Wien Klin Wochenschr Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Áustria

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Fluorescência / Ensaio de Imunoadsorção Enzimática / Herpesvirus Humano 4 / Infecções por Vírus Epstein-Barr / Testes Diagnósticos de Rotina / Anticorpos Antivirais Tipo de estudo: Diagnostic_studies / Guideline / Observational_studies / Qualitative_research Limite: Humans Idioma: En Revista: Wien Klin Wochenschr Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Áustria