Your browser doesn't support javascript.
loading
Note: Operation of gamma-ray microcalorimeters at elevated count rates using filters with constraints.
Alpert, B K; Horansky, R D; Bennett, D A; Doriese, W B; Fowler, J W; Hoover, A S; Rabin, M W; Ullom, J N.
Afiliação
  • Alpert BK; National Institute of Standards and Technology, Boulder, Colorado 80305, USA.
Rev Sci Instrum ; 84(5): 056107, 2013 May.
Article em En | MEDLINE | ID: mdl-23742605
ABSTRACT
Microcalorimeter sensors operated near 0.1 K can measure the energy of individual x- and gamma-ray photons with significantly more precision than conventional semiconductor technologies. Both microcalorimeter arrays and higher per pixel count rates are desirable to increase the total throughput of spectrometers based on these devices. The millisecond recovery time of gamma-ray microcalorimeters and the resulting pulse pileup are significant obstacles to high per pixel count rates. Here, we demonstrate operation of a microcalorimeter detector at elevated count rates by use of convolution filters designed to be orthogonal to the exponential tail of a preceding pulse. These filters allow operation at 50% higher count rates than conventional filters while largely preserving sensor energy resolution.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Estados Unidos