Mechanical evaluation of articulating instruments and cross-handed manipulation in laparoendoscopic single-site surgery.
Xu, An An; Zhu, Jiang Fan; Xie, Xiaofeng; Su, Yuantao.
Afiliação
  • Xu AA; East Hospital of Tongji University, Shanghai, China.
  • Zhu JF; East Hospital of Tongji University, Shanghai, China zhujiangfan@hotmail.com.
  • Xie X; East Hospital of Tongji University, Shanghai, China.
  • Su Y; East Hospital of Tongji University, Shanghai, China.
Surg Innov ; 21(4): 398-402, 2014 Aug.
Article em En | MEDLINE | ID: mdl-24270604
ABSTRACT

BACKGROUND:

Laparoendoscopic single-site surgery (LESS) is limited by loss of triangulation and internal instruments conflict. To overcome these difficulties, some concepts have been introduced, namely, articulating instruments and cross-handed manipulation, which causes the right hand to control the left instrument tip and vice versa. The aim of this study was to compare task performance with different approaches based on a mechanical evaluation platform.

METHODS:

A LESS mechanical evaluation platform was set up to investigate the performance of 2 tasks (suture pass-through rings and clip-cut) with 3 different settings uncrossed manipulation with straight instruments (group A, the control group), uncrossed manipulation with articulating instruments (group B), and cross-handed manipulation with articulating instruments (group C). The operation time and average load required for accomplishment of the standard tasks were measured.

RESULTS:

Group A presented significantly better time scores than group B, and group C consumed the longest time to accomplish the 2 tasks (P < .05). Comparing of average load required to perform the suture pass-through rings task, it differed significantly between dominant and nondominant hand in all groups (P < .01) and was less in group A and group B than group C in dominant hand (P < .01), while it was almost the same in all groups in the nondominant hand. In terms of average load requirement to accomplish clip-cut task, it was almost equal not only between group A and B but also between dominant and nondominant hand while the increase reached statistical significance when comparing group C with other groups (P < .05).

CONCLUSIONS:

Compared with conventional devices and maneuvering techniques, articulating instruments and cross-handed manipulation are associated with longer operation time and higher workload. Instruments with better maneuverability should be developed in the future for LESS.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Instrumentos Cirúrgicos / Laparoscopia / Laparoscópios / Modelos Anatômicos Tipo de estudo: Evaluation_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Surg Innov Ano de publicação: 2014 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Instrumentos Cirúrgicos / Laparoscopia / Laparoscópios / Modelos Anatômicos Tipo de estudo: Evaluation_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Surg Innov Ano de publicação: 2014 Tipo de documento: Article País de afiliação: China