Mass spectrometric analysis of anatoxin-a.
Ross, M M; Kidwell, D A; Callahan, J H.
Afiliação
  • Ross MM; Chemistry Division, Naval Research Laboratory, Washington, DC 20375-5000.
J Anal Toxicol ; 13(6): 317-21, 1989.
Article em En | MEDLINE | ID: mdl-2578029
ABSTRACT
Secondary ion mass spectrometry (SIMS), gas chromatography/mass spectrometry (GC/MS), desorption chemical ionization/mass spectrometry (DCI/MS), and thermospray/mass spectrometry (TSP/MS) were evaluated for the detection of a low-molecular weight biological toxin, anatoxin-a. The lowest detection limit was obtained using isobutane or ammonia DCI/MS. Tandem mass spectrometry (MS/MS) was used to enhance the selectivity and sensitivity of the analysis. The detection limit for the pure material is 10 pg, and 100 pg of the toxin can be analyzed directly from urine using DCI/MS/MS.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas / Toxinas Bacterianas / Toxinas Marinhas Tipo de estudo: Diagnostic_studies / Evaluation_studies Idioma: En Revista: J Anal Toxicol Ano de publicação: 1989 Tipo de documento: Article