Improved MRI thermometry with multiple-echo spirals.
Marx, Michael; Butts Pauly, Kim.
Afiliação
  • Marx M; Department of Radiology, Stanford University, Stanford, California, USA.
  • Butts Pauly K; Department of Radiology, Stanford University, Stanford, California, USA.
Magn Reson Med ; 76(3): 747-56, 2016 09.
Article em En | MEDLINE | ID: mdl-26332512
ABSTRACT

PURPOSE:

Low-bandwidth PRF shift thermometry is used to guide HIFU ablation treatments. Low sampling bandwidth is needed for high signal-to-noise ratio with short acquisition times, but can lead to off-resonance artifacts. In this work, improved multiple-echo thermometry is presented that allows for high bandwidth and reduced artifacts. It is also demonstrated with spiral sampling, to improve the trade-off between resolution, speed, and measurement precision.

METHODS:

Four multiple-echo thermometry sequences were tested in vivo, one using two-dimensional Fourier transform (2DFT) sampling and three using spirals. The spiral sequences were individually optimized for resolution, for speed, and for precision. Multifrequency reconstruction was used to correct for off-resonance spiral artifacts. Additionally, two different multiecho temperature reconstructions were compared.

RESULTS:

Weighted combination of per-echo phase differences gave significantly better precision than least squares off-resonance estimation. Multiple-echo 2DFT sequence obtained precision similar to single-echo 2DFT, while greatly increasing sampling bandwidth. The multiecho spiral acquisitions achieved 2× better resolution, 2.9× better uncertainty, or 3.4× faster acquisition time, without negatively impacting the other two design parameters as compared to single-echo 2DFT.

CONCLUSION:

Multiecho spiral thermometry greatly improves the capabilities of temperature monitoring, and could improve transcranial treatment monitoring capabilities. Magn Reson Med 76747-756, 2016. © 2015 Wiley Periodicals, Inc.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Processamento de Sinais Assistido por Computador / Encéfalo / Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Termografia Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Algoritmos / Processamento de Sinais Assistido por Computador / Encéfalo / Imageamento por Ressonância Magnética / Interpretação de Imagem Assistida por Computador / Aumento da Imagem / Termografia Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Estados Unidos