Your browser doesn't support javascript.
loading
Low contrast medium-volume third-generation dual-source computed tomography angiography for transcatheter aortic valve replacement planning.
Felmly, Lloyd M; De Cecco, Carlo N; Schoepf, U Joseph; Varga-Szemes, Akos; Mangold, Stefanie; McQuiston, Andrew D; Litwin, Sheldon E; Bayer, Richard R; Vogl, Thomas J; Wichmann, Julian L.
Afiliação
  • Felmly LM; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA.
  • De Cecco CN; Division of Cardiothoracic Surgery, Department of Surgery, Medical University of South Carolina, Charleston, SC, USA.
  • Schoepf UJ; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA.
  • Varga-Szemes A; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA. schoepf@musc.edu.
  • Mangold S; Division of Cardiology, Department of Medicine, Medical University of South Carolina, Charleston, SC, USA. schoepf@musc.edu.
  • McQuiston AD; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA.
  • Litwin SE; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA.
  • Bayer RR; Department of Diagnostic and Interventional Radiology, University Hospital of Tübingen, Tübingen, Germany.
  • Vogl TJ; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA.
  • Wichmann JL; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA.
Eur Radiol ; 27(5): 1944-1953, 2017 May.
Article em En | MEDLINE | ID: mdl-27553939
ABSTRACT

PURPOSE:

To investigate feasibility, image quality and safety of low-tube-voltage, low-contrast-volume comprehensive cardiac and aortoiliac CT angiography (CTA) for planning transcatheter aortic valve replacement (TAVR). MATERIALS AND

METHODS:

Forty consecutive TAVR candidates prospectively underwent combined CTA of the aortic root and vascular access route (270 mgI/ml iodixanol). Patients were assigned to group A (second-generation dual-source CT [DSCT], 100 kV, 60 ml contrast, 4.0 ml/s flow rate) or group B (third-generation DSCT, 70 kV, 40 ml contrast, 2.5 ml/s flow rate). Vascular attenuation, noise, signal-to-noise (SNR) and contrast-to-noise ratios (CNR) were compared. Subjective image quality was assessed by two observers. Estimated glomerular filtration (eGFR) at CTA and follow-up were measured.

RESULTS:

Besides a higher body-mass-index in group B (24.8±3.8 kg/m2 vs. 28.1±5.4 kg/m2, P=0.0339), patient characteristics between groups were similar (P≥0.0922). Aortoiliac SNR (P=0.0003) was higher in group B. Cardiac SNR (P=0.0003) and CNR (P=0.0181) were higher in group A. Subjective image quality was similar (P≥0.213) except for aortoiliac image noise (4.42 vs. 4.12, P=0.0374). TAVR-planning measurements were successfully obtained in all patients. There were no significant changes in eGFR among and between groups during follow-up (P≥0.302).

CONCLUSION:

TAVR candidates can be safely and effectively evaluated by a comprehensive CTA protocol with low contrast volume using low-tube-voltage acquisition. KEY POINTS • Third-generation dual-source CT facilitates low-tube-voltage acquisition. • TAVR planning can be performed with reduced contrast volume and radiation dose. • TAVR-planning CT did not result in changes in creatinine levels at follow-up. • TAVR candidates can be safely evaluated by comprehensive low-tube-voltage CT angiography.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Substituição da Valva Aórtica Transcateter / Angiografia por Tomografia Computadorizada Tipo de estudo: Evaluation_studies / Observational_studies Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Substituição da Valva Aórtica Transcateter / Angiografia por Tomografia Computadorizada Tipo de estudo: Evaluation_studies / Observational_studies Limite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos