Physiologic musculofascial compliance following reinforcement with electrospun polycaprolactone-ureidopyrimidinone mesh in a rat model.
Hympanova, Lucie; Mori da Cunha, Marina Gabriela Monteiro Carvalho; Rynkevic, Rita; Zündel, Manuel; Gallego, Monica Ramos; Vange, Jakob; Callewaert, Geertje; Urbankova, Iva; Van der Aa, Frank; Mazza, Edoardo; Deprest, Jan.
Afiliação
  • Hympanova L; Centre for Surgical Technologies, KU Leuven, Leuven, Belgium; Department of Development and Regeneration, KU Leuven, Leuven, Belgium; Institute for the Care of Mother and Child, Charles University, Prague, Czech Republic. Electronic address: lucie.hympanova@upmd.eu.
  • Mori da Cunha MGMC; Centre for Surgical Technologies, KU Leuven, Leuven, Belgium; Department of Development and Regeneration, KU Leuven, Leuven, Belgium. Electronic address: biamori@gmail.com.
  • Rynkevic R; Centre for Surgical Technologies, KU Leuven, Leuven, Belgium; Department of Development and Regeneration, KU Leuven, Leuven, Belgium; INEGI, Faculdade de Engenharia da Universidade do Porto, Porto, Portugal. Electronic address: r.rynkevic@gmail.com.
  • Zündel M; Institute of Mechanical Systems, ETH Zurich, Zurich, Switzerland. Electronic address: zuendel@imes.mavt.ethz.ch.
  • Gallego MR; Coloplast A/S, Global R&D, Biomaterials, Holtedam, Humlebæk, Denmark. Electronic address: dkmrg@coloplast.com.
  • Vange J; Coloplast A/S, Global R&D, Biomaterials, Holtedam, Humlebæk, Denmark. Electronic address: dkjv@coloplast.com.
  • Callewaert G; Centre for Surgical Technologies, KU Leuven, Leuven, Belgium; Department of Development and Regeneration, KU Leuven, Leuven, Belgium; Pelvic Floor Unit, University Hospitals KU Leuven, Leuven, Belgium. Electronic address: gjecallewaert@gmail.com.
  • Urbankova I; Centre for Surgical Technologies, KU Leuven, Leuven, Belgium; Department of Development and Regeneration, KU Leuven, Leuven, Belgium; Institute for the Care of Mother and Child, Charles University, Prague, Czech Republic. Electronic address: iva.urbankova@upmd.eu.
  • Van der Aa F; Department of Development and Regeneration, KU Leuven, Leuven, Belgium; Department of Urology, University Hospitals Leuven, Leuven, Belgium. Electronic address: frank.vanderaa@kuleuven.be.
  • Mazza E; Institute of Mechanical Systems, ETH Zurich, Zurich, Switzerland; EMPA, Swiss Federal Laboratories for Materials Science and Technology, Dübendorf, Switzerland. Electronic address: mazza@imes.mavt.ethz.ch.
  • Deprest J; Centre for Surgical Technologies, KU Leuven, Leuven, Belgium; Department of Development and Regeneration, KU Leuven, Leuven, Belgium; Pelvic Floor Unit, University Hospitals KU Leuven, Leuven, Belgium. Electronic address: Jan.Deprest@uzleuven.be.
J Mech Behav Biomed Mater ; 74: 349-357, 2017 10.
Article em En | MEDLINE | ID: mdl-28668592
PURPOSE: Electrospun meshes may be considered as substitutes to textile polypropylene implants. We compared the host response and biomechanical properties of the rat abdominal wall following reinforcement with either polycaprolactone (PCL) modified with ureidopyrimidinone-motifs (UPy) or polypropylene mesh. METHODS: First we measured the response to cyclic uniaxial load within the physiological range both dry (room temperature) and wet (body temperature). 36 rats underwent primary repair of a full-thickness abdominal wall defect with a polypropylene suture (native tissue repair), or reinforced with either UPy-PCL or ultra-light weight polypropylene mesh (n = 12/group). Sacrifice was at 7 and 42 days. Outcomes were compliance of explants, mesh dimensions, graft related complications and semi-quantitative assessment of inflammatory cell (sub) types, neovascularization and remodeling. RESULTS: Dry UPy-PCL implants are less stiff than polypropylene, both are more compliant in wet conditions. Polypropylene loses stiffness on cyclic loading. Both implant types were well incorporated without clinically obvious degradation or herniation. Exposure rates were similar (n = 2/12) as well as mesh contraction. There was no reinforcement at low loads, while, at higher tension, polypropylene explants were much stiffer than UPy-PCL. The latter was initially weaker yet by 42 days it had a compliance similar to native abdominal wall. There were eventually more foreign body giant cells around UPy-PCL fibers yet the amount of M1 subtype macrophages was higher than in polypropylene explants. There were less neovascularization and collagen deposition. CONCLUSION: Abdominal wall reconstruction with electrospun UPy-PCL mesh does not compromise physiologic tissue biomechanical properties, yet provokes a vivid inflammatory reaction.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Telas Cirúrgicas / Músculo Esquelético / Parede Abdominal / Fáscia Limite: Animals Idioma: En Revista: J Mech Behav Biomed Mater Assunto da revista: ENGENHARIA BIOMEDICA Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Telas Cirúrgicas / Músculo Esquelético / Parede Abdominal / Fáscia Limite: Animals Idioma: En Revista: J Mech Behav Biomed Mater Assunto da revista: ENGENHARIA BIOMEDICA Ano de publicação: 2017 Tipo de documento: Article